Vertical streaks
White vertical streaks
Hue defect
<User Check>
- Clean the CIS glass.
- Clean the feeding path.
Step
Cause
Remedy
1
First side or second side CIS unit
failure
Replace the first side or second side CIS
unit.
<User Check>
- Wipe the first and second side CIS glass four times in the same
direction with a cleaning cloth.
- Clean the feeding path.
Step Cause
Remedy
1
First side or second side CIS unit
failure
Replace the first side or second side CIS
unit.
<User Check>
- Clean the document glass or CIS glass.
Step Cause
Remedy
1
Color parameter failure
Execute the maintenance mode function
"Acquire white level data and set CIS scan
area". (Refer to
.)
2
Dirty calibration sheet
Replace the calibration sheet with a new
one.
3
First side or second side CIS unit
failure
Replace the first side or second side CIS
unit.
4
Main PCB failure
Replace the main PCB ASSY.