10.3

Start Auto-Zero

With no flow, use the push-button switch (#) on the

mini CORI-FLOW™

to start the zero adjustment

procedure (LEDs first will go off). (1), (2) and (3):
Press the push-button (#) and hold it for 8..12
seconds (each 4 seconds the LED-indication will
change). (4): Release the push-button (#) when the
green LED is on.

10.4

Zeroing procedure

The zeroing procedure will start now and the green
LED will blink fast. The procedure will take at least
45 sec. When the signal is not stable the procedure
will take longer (max. 6 retries will be performed).
The red LED will blink when signal is unstable/noisy.
Make sure that there is no flow through the

mini

CORI-FLOW™

when performing the zeroing

procedure and avoid vibrations !

45 sec.

10.5

Ready

When indication is showing 0% signal and the
green indication LED is continuously on, then zero
has been performed well. Instrument is in normal
operation mode now and ready for use.

Measurement check

For a controller: send a setpoint to the

mini CORI-FLOW™

and check the measured value.

Make sure the

mini CORI-FLOW™

indicates 0% at zero flow.

Your

mini CORI-FLOW™

is now ready for operation.

Bronkhorst High-Tech B.V., Nijverheidsstraat 1a, NL-7261AK Ruurlo The Netherlands

T

+31(0)573 45 88 00

F

+31(0)573 45 88 08

I

www.bronkhorst.com

E

[email protected]

Quick Installation Guide

Doc nr.: 9.17.052H   Date: 10-12-2015

Содержание mini CORI-FLOW

Страница 1: ...Quick Installation Guide Doc nr 9 17 052H Date 10 12 2015 ...

Страница 2: ...Plot Manual document nr 9 17 030 Depending on optional fieldbus interface Fieldbus interface description Manual Hookup diagram RS232 Interface with FLOW BUS protocol 9 17 027 9 16 044 FLOW BUS Interface 9 17 024 9 16 048 PROFIBUS DP Interface 9 17 025 9 16 049 DeviceNet Interface 9 17 026 9 16 050 Modbus interface 9 17 035 9 16 066 i www These documents can be downloaded from the website https www...

Страница 3: ... water particles may cause undesired reactions such as corrosion 4 1 Mount install instrument properly Install the mini CORI FLOW Meter Controller in the line and tighten the fittings according to the instructions of the supplier of the fittings Mount the mini CORI FLOW instrument with screws in the body to a rigid stiff base body or heavy mass such as a wall heavy rig or stable steel construction...

Страница 4: ...leaks before applying fluid pressure Especially if toxic explosive or other dangerous fluids are used Gas tight is not the same as liquid tight Please note that connections which are tight for liquid could still be untight for gas This can result in Air enclosure in the liquid e g when using pumps which might lead to errors in measurement 6 Electrical connection Electrical connections must be made...

Страница 5: ...Special control mode for purging In case of a mini CORI FLOW controller give setpoint 100 to control the valve or the pump It is also possible to use special control mode 8 to fully open the valve or set the pump at max rpm using a digital interface This will bypass the PID controller and might be usefull when having the mini CORI FLOW set to a low capacity It will ensure you to get the highest po...

Страница 6: ...The red LED will blink when signal is unstable noisy Make sure that there is no flow through the mini CORI FLOW when performing the zeroing procedure and avoid vibrations 45 sec 10 5 Ready When indication is showing 0 signal and the green indication LED is continuously on then zero has been performed well Instrument is in normal operation mode now and ready for use Measurement check For a controll...

Отзывы: