Instruction Manual ES-FLOW™
9.17.116B
34
Fsetpoint
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
0…3.4E+38
206
33/3
0xA119…0xA11A/ 41241…41242
Floating point variant of
Setpoint
.
Fsetpoint
shows the setpoint in the capacity and capacity unit for which the instrument
has been set.
Fsetpoint
is dependent of
Capacity Unit
and
Sensor Type
.
Setpoint Slope
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned int
RW
0…30000
10
1/2
0x0022/35
The value of this parameter represents the time it would take to adjust the setpoint if it were changed from 0 to 100%. This
feature can be used to smooth 'nervous' controller behavior, e.g. to reduce setpoint overshoot or undershoot.
The supported range corresponds with 0…3000 seconds. Default value = 0.
Example:
If
Setpoint Slope
= 100 it will take 10 seconds to adjust the setpoint if it is changed from 0 to 100%. A setpoint change of 20%
will take (20%/100%)*10 seconds = 2 seconds.
Analog Input
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned int
R
0…65535
11
1/3
0x0023/36
This parameter contains a digital translation of the analog input signal (if applicable).
Valve Output
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned long
RW
0…
16777215
55
114/1
0xF208…0xF209/61961…61962
Digital steering signal for driving the control valve, where 0…16777215 corresponds with 0…100%.
3.4.4
Device identification
User Tag
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[16]
RW
-
115
113/6
0xF130…0xF137/ 61745…61752
With this parameter, the instrument can be given a custom tag name, with a maximum of 16 characters.
Customer Model
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[16]
RW
-
93
113/4
0xF120…0xF127/ 61729…61736
This parameter is used to add extra information to the model number information, such as a customer-specific model
number.
Serial Number
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[20]
R
-
92
113/3
0xF118…0xF11F/ 61721…61728
Instrument serial number for identification.