The
Master slave
item is only available if one of the slave control modes is selected. Availability of
Master node
and
Slave factor
depends on the selected mode:
Control mode
Master node
Slave factor
fb slave
analog slave
fb ana slave
See also
for more information about setting up a master/slave relationship between
instruments.
3.3.3.3 Counter
Bronkhorst® flow meters have a built-in counter function, which can be used to monitor and/or control the amount of
media flowing through the instrument. The flow can be stopped or changed when a certain limit is reached. Until the
counter is reset, the counter setpoint overrules the regular setpoint.
In the
Counter
sub menu, the following parameters can be edited:
Parameter
Description
Supported values
Mode
Selected counter mode
·
off
·
up to limit
·
up
Reset
Reset method
·
automatic
·
manual
Limit
Counter limit or batch size
Setpoint range
Setpoint change
Specifies whether or not to change the setpoint after
reaching the counter limit
·
yes
·
no
Setpoint
New setpoint after reaching counter limit (until counter
reset)
Setpoint range
3.3.3.4 Alarm