Symptom

Possible cause

Action

Inlet pressure or differential

pressure out of bounds

Set inlet pressure to a value within

specifications

Measured value rises, but never

reaches setpoint

Piping, filters and/or control valve

clogged or blocked

·

Flush fluidic system with clean, dry air. If

problem persists, contact Bronkhorst.

·

For external proportional control valves:

supply 0…15 Vdc and operational inlet

pressure to valve and slowly increase

voltage. If valve does not open, clean parts

and re-adjust valve

Inlet pressure too low

Increase inlet pressure

Outlet pressure too high

Check/decrease outlet pressure

Process outlet blocked

Check process outlet and downstream piping

Measured value or output signal

(much) lower than setpoint

Inlet pressure or differential

pressure too low

·

Increase inlet pressure

·

Use instrument in conditions it was designed

for

Process gas condensation

Decrease inlet pressure or increase gas

temperature

·

Piping, filters and/or control

valve clogged or blocked

·

Sensor blocked or contaminated

Flush fluidic system with clean, dry air. If

problem persists, contact Bronkhorst.

Supplied fluid type does not match

configured fluid type

Supply instrument with fluid it was configured

for

Measured value or output signal

indicates a flow, while there is

none

Mounting orientation and/or

ambient conditions changed

significantly

·

Use instrument in conditions it was designed

for

·

Adjust zero point (see 

Zeroing before use

)

System leakage

Check the fluidic system for leakage. Follow

vendor instructions when installing third party

components (e.g. adapters, tubing, valves)

Continuous maximum measured

value or output signal

Inlet pressure too high

Check inlet pressure

Valve fully open

Close valve

Sensor failure

Return equipment to factory

Содержание EL-FLOW Classic Series

Страница 1: ...Pleaseread this Instruction Manualcarefully beforeinstalling and operating theinstrument Notfollowing theguidelines could resultin personalinjury and ordamageto theequipment EL FLOW Classic series Thermal Mass Flow Controllers Instruction Manual ...

Страница 2: ...mationwill facilitatetheuseoftheinstrument and or contributetoitsoptimal performance Additional infoavailableontheinternet or from your local salesrepresentative Receipt of equipment Check the outside package box for damage incurred during shipment If the box is damaged then the local carrier must be notified at once regarding his liability if so required At the same time a report should be submit...

Страница 3: ...intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury Read the operating information carefully before using the product Before operating make sure the line cord is connected to a properly grounded power receptacle Inspect the connecting cables for cracks or breaks before each use The equipment and accessor...

Страница 4: ...Bronkhorst Instruction Manual EL FLOW Classic series 9 17 085D 4 ...

Страница 5: ...2 6 2 9 2 7 Powering up and powering down 9 2 8 Purging before use 10 2 9 Zeroing before use 10 2 10 Calibration 11 2 11 Model key 12 Basic operation 3 12 3 1 General 12 3 2 LED indications 13 3 3 Multifunctional switch 13 Normal operating functions 3 3 1 13 Power up functions 3 3 2 14 Advanced operation 4 14 4 1 Using other gases than specified 14 4 2 Conversion factor calculation 14 4 3 Maximum ...

Страница 6: ...hmediaandconditions e g pressure temperature asspecifiedat orderingtime Ifyouareplanningtousetheproduct includinganythirdpartycomponents suppliedbyBronkhorst suchaspumpsor valves withother mediaand or other conditions alwaysbesuretocheckthe wettedmaterials includingseals for compatibility Seethetechnical specificationsoftheproduct andconsult thirdparty documentation ifapplicable tochecktheincorpor...

Страница 7: ...ed with a clean dry and inert gas In case of severe contamination cleaning the inside of the device may be required After cleaning recalibration of the instrument is recommended Inexpertlyservicinginstrumentscanleadtoseriouspersonal injuryand or damagetotheinstrument or thesystem it is usedin Therefore servicingmust beperformedbytrainedandqualifiedpersonnel Contact your local Bronkhorst representa...

Страница 8: ...g The bottom side of the EL FLOW Classic is fitted with two threaded mounting holes for stable mechanical fixation of the instrument Refer to the dimensional drawing for the exact position of the mounting holes For optimal performance observethefollowingguidelines Preferably mount theEL FLOW Classic inanupright position especiallyiftheoperatingpressureishigher than10 bar Whenmountingtheinstrument ...

Страница 9: ...ling proceduresmust befollowedduringinstallation dis connectingandremovingtheelectronics Thedevicedescribedinthismanual carriestheCE markandiscompliant with the concerning EMC requirements However compliancewiththeEMC requirementsisnot possiblewithout theuseofproper cablesandconnector gland assemblies Bronkhorst recommendstheuseoftheir standardcables Thesecableshavetheright connectorsandifloose en...

Страница 10: ...ing might take longer and the nearest point to zero is accepted The procedure will take approximately 10 seconds When the indication is showing 0 signal and the green indication LED is glowing continuously again then the zeroing action was successful 2 10 Calibration EL FLOW Classicinstruments are Air or N2 calibrated Bronkhorst High Tech B V certifies that all instruments meet the rated accuracy ...

Страница 11: ...Bronkhorst Instruction Manual EL FLOW Classic series 9 17 085D 11 2 11 Model key Example F 201CL 009 1K0 A Gas CO Range 1 SLM Analog output 0 5 Vdc ...

Страница 12: ...ndon theserial number label 3 2 LED indications green Mode operation mode indication red Error error warning messages The tables below list the different LED indications Green Pattern Time Indication on continuous Normal operation mode off continuous Power off or program not running blink 0 2 sec on 0 2 sec off Special function mode the instrument is busy performing a special function e g auto zer...

Страница 13: ...error messages and restart the instrument on off 8 12 sec Auto zero re adjust the zero point of the instrument flow meters controllers on on 12 16 sec Factory mode do not use SeeZeroingbeforeusefor backgroundinformationandinstructionsonhowtoadjust thezeropoint ofaninstrument Never perform azeroingprocedurebeforehavingtakennoticeoftheinstructions 3 3 2 Power up functions In order to access these fu...

Страница 14: ...abasecalledFLUIDAT Thisdatabasealso providesacollectionofroutinestocalculatephysical propertiesofgasesandliquids not onlyat 20 C 1 atm but at any temperature pressurecombination bothfor gasesandfor liquids CheckFLUIDAT ontheNet onhttp www fluidat com Consult your distributor for moredetails Also see section Calibration 4 3 Maximum pressure drop For pilot solenoid operated control valves with small...

Страница 15: ...ection Check cable hook up No setpoint given or setpoint too low Inlet pressure or differential pressure too low Increase inlet pressure Piping filters and or control valve clogged or blocked Flush fluidic system with clean dry air If problem persists contact Bronkhorst For external proportional control valves supply 0 15 Vdc and operational inlet pressure to valve and slowly increase voltage If v...

Страница 16: ...sure or differential pressure too low Increase inlet pressure Use instrument in conditions it was designed for Process gas condensation Decrease inlet pressure or increase gas temperature Piping filters and or control valve clogged or blocked Sensor blocked or contaminated Flush fluidic system with clean dry air If problem persists contact Bronkhorst Supplied fluid type does not match configured f...

Страница 17: ...ight product for your application Contact sales by e mail sales bronkhorst com For after sales questions our Customer Service Department is available with help and guidance To contact CSD by e mail support bronkhorst com No matter the time zone our experts within the Support Group are available to answer your request immediately or ensure appropriate further action Our experts can be reached at 31...

Страница 18: ...edfrom theService Support sectionoftheBronkhorst website www bronkhorst com Important Clearly note on top of the package the customs clearance number of Bronkhorst High Tech B V NL801989978B01 only if applicable otherwise contact your local Bronkhorst representative for local arrangements 6 2 Disposal end of lifetime Within the scope of the European Union manufacturers of electrical and electronic...

Отзывы: