Bronkhorst®

Instruction Manual EL-FLOW® Classic series

9.17.085D

13

3.3

Multifunctional switch

Some special functions of the instrument can be started manually using the multifunctional switch near the indication

LEDs.

3.3.1

Normal operating functions

·

In order to access these functions, press and hold the switch while the instrument is in normal operation mode (green

LED glowing).

·

As long as the switch is held, the LEDs show a repeating sequence of patterns, where each pattern indicates a function.

·

All indications in this sequence are continuous.

·

Each pattern is shown for a number of seconds; in the table below the column 

Hold time

 indicates the time frame within

the sequence where the LEDs show the associated pattern.

·

To start a function, release the switch when the LEDs show the pattern of the required function.

(green)

(red)

Hold time Function

off

off

0…4 sec

No action

off

on

4…8 sec

Reset instrument; clear all warnings and error messages and restart the instrument

on

off

8…12 sec Auto-zero; re-adjust the zero-point of the instrument (flow meters/controllers)

on

on

12…16 sec Factory mode, do not use

See 

Zeroing before use

 for background information and instructions on how to adjust the zero point of an instrument.

Never perform a zeroing procedure before having taken notice of the instructions.

3.3.2

Power-up functions

·

In order to access these functions, press and hold the switch while powering up the instrument.

·

As long as the switch is held, the LEDs show a repeating sequence of patterns, where each pattern indicates a function.

·

All indications in this sequence are flashing (0.2 sec on, 0.2 sec off).

·

Each pattern is shown for a number of seconds; in the table below the column 

Hold time

 indicates the time frame within

the sequence where the LEDs show the associated pattern.

·

To start a function, release the switch when the LEDs show the pattern of the required function.

(green)

(red)

Hold time Function

off

off

0…4 sec

No action

off

on

4…8 sec

Restore factory settings (All parameter settings will be restored)

on

off

8…12 sec No action

on

on

12…16 sec Factory mode, do not use

Содержание EL-FLOW Classic Series

Страница 1: ...Pleaseread this Instruction Manualcarefully beforeinstalling and operating theinstrument Notfollowing theguidelines could resultin personalinjury and ordamageto theequipment EL FLOW Classic series Thermal Mass Flow Controllers Instruction Manual ...

Страница 2: ...mationwill facilitatetheuseoftheinstrument and or contributetoitsoptimal performance Additional infoavailableontheinternet or from your local salesrepresentative Receipt of equipment Check the outside package box for damage incurred during shipment If the box is damaged then the local carrier must be notified at once regarding his liability if so required At the same time a report should be submit...

Страница 3: ...intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury Read the operating information carefully before using the product Before operating make sure the line cord is connected to a properly grounded power receptacle Inspect the connecting cables for cracks or breaks before each use The equipment and accessor...

Страница 4: ...Bronkhorst Instruction Manual EL FLOW Classic series 9 17 085D 4 ...

Страница 5: ...2 6 2 9 2 7 Powering up and powering down 9 2 8 Purging before use 10 2 9 Zeroing before use 10 2 10 Calibration 11 2 11 Model key 12 Basic operation 3 12 3 1 General 12 3 2 LED indications 13 3 3 Multifunctional switch 13 Normal operating functions 3 3 1 13 Power up functions 3 3 2 14 Advanced operation 4 14 4 1 Using other gases than specified 14 4 2 Conversion factor calculation 14 4 3 Maximum ...

Страница 6: ...hmediaandconditions e g pressure temperature asspecifiedat orderingtime Ifyouareplanningtousetheproduct includinganythirdpartycomponents suppliedbyBronkhorst suchaspumpsor valves withother mediaand or other conditions alwaysbesuretocheckthe wettedmaterials includingseals for compatibility Seethetechnical specificationsoftheproduct andconsult thirdparty documentation ifapplicable tochecktheincorpor...

Страница 7: ...ed with a clean dry and inert gas In case of severe contamination cleaning the inside of the device may be required After cleaning recalibration of the instrument is recommended Inexpertlyservicinginstrumentscanleadtoseriouspersonal injuryand or damagetotheinstrument or thesystem it is usedin Therefore servicingmust beperformedbytrainedandqualifiedpersonnel Contact your local Bronkhorst representa...

Страница 8: ...g The bottom side of the EL FLOW Classic is fitted with two threaded mounting holes for stable mechanical fixation of the instrument Refer to the dimensional drawing for the exact position of the mounting holes For optimal performance observethefollowingguidelines Preferably mount theEL FLOW Classic inanupright position especiallyiftheoperatingpressureishigher than10 bar Whenmountingtheinstrument ...

Страница 9: ...ling proceduresmust befollowedduringinstallation dis connectingandremovingtheelectronics Thedevicedescribedinthismanual carriestheCE markandiscompliant with the concerning EMC requirements However compliancewiththeEMC requirementsisnot possiblewithout theuseofproper cablesandconnector gland assemblies Bronkhorst recommendstheuseoftheir standardcables Thesecableshavetheright connectorsandifloose en...

Страница 10: ...ing might take longer and the nearest point to zero is accepted The procedure will take approximately 10 seconds When the indication is showing 0 signal and the green indication LED is glowing continuously again then the zeroing action was successful 2 10 Calibration EL FLOW Classicinstruments are Air or N2 calibrated Bronkhorst High Tech B V certifies that all instruments meet the rated accuracy ...

Страница 11: ...Bronkhorst Instruction Manual EL FLOW Classic series 9 17 085D 11 2 11 Model key Example F 201CL 009 1K0 A Gas CO Range 1 SLM Analog output 0 5 Vdc ...

Страница 12: ...ndon theserial number label 3 2 LED indications green Mode operation mode indication red Error error warning messages The tables below list the different LED indications Green Pattern Time Indication on continuous Normal operation mode off continuous Power off or program not running blink 0 2 sec on 0 2 sec off Special function mode the instrument is busy performing a special function e g auto zer...

Страница 13: ...error messages and restart the instrument on off 8 12 sec Auto zero re adjust the zero point of the instrument flow meters controllers on on 12 16 sec Factory mode do not use SeeZeroingbeforeusefor backgroundinformationandinstructionsonhowtoadjust thezeropoint ofaninstrument Never perform azeroingprocedurebeforehavingtakennoticeoftheinstructions 3 3 2 Power up functions In order to access these fu...

Страница 14: ...abasecalledFLUIDAT Thisdatabasealso providesacollectionofroutinestocalculatephysical propertiesofgasesandliquids not onlyat 20 C 1 atm but at any temperature pressurecombination bothfor gasesandfor liquids CheckFLUIDAT ontheNet onhttp www fluidat com Consult your distributor for moredetails Also see section Calibration 4 3 Maximum pressure drop For pilot solenoid operated control valves with small...

Страница 15: ...ection Check cable hook up No setpoint given or setpoint too low Inlet pressure or differential pressure too low Increase inlet pressure Piping filters and or control valve clogged or blocked Flush fluidic system with clean dry air If problem persists contact Bronkhorst For external proportional control valves supply 0 15 Vdc and operational inlet pressure to valve and slowly increase voltage If v...

Страница 16: ...sure or differential pressure too low Increase inlet pressure Use instrument in conditions it was designed for Process gas condensation Decrease inlet pressure or increase gas temperature Piping filters and or control valve clogged or blocked Sensor blocked or contaminated Flush fluidic system with clean dry air If problem persists contact Bronkhorst Supplied fluid type does not match configured f...

Страница 17: ...ight product for your application Contact sales by e mail sales bronkhorst com For after sales questions our Customer Service Department is available with help and guidance To contact CSD by e mail support bronkhorst com No matter the time zone our experts within the Support Group are available to answer your request immediately or ensure appropriate further action Our experts can be reached at 31...

Страница 18: ...edfrom theService Support sectionoftheBronkhorst website www bronkhorst com Important Clearly note on top of the package the customs clearance number of Bronkhorst High Tech B V NL801989978B01 only if applicable otherwise contact your local Bronkhorst representative for local arrangements 6 2 Disposal end of lifetime Within the scope of the European Union manufacturers of electrical and electronic...

Отзывы: