MODBUS REGISTERS
PARAMETER
NAME
PARAMETER
TYPE
ACCES
S
PDU ADDRESS
REGISTER NUMBER
REMARK
Hex
Dec
Hex
Dec
Open from zero c. resp.
Unsigned char
RW
0x0E52
3666
0x0E53
3667
Calibration mode
Unsigned char
RW
0x0E61
3681
0x0E62
3682
Monitor mode
Unsigned char
RW
0x0E62
3682
0x0E63
3683
Reset
Unsigned char
W
0x0E68
3688
0x0E69
3689
Fieldbus2 address
Unsigned char
RW
0x0F8A
3978
0x0F8B
3979
Modbus address
OEMD2
Fieldbus2 parity
Unsigned char
RW
0x0F8C
3980
0x0F8D
3981
Modbus parity
OEMD2
Fieldbus1 address
Unsigned char
RW
0x0FAA
4010
0x0FAB
4011
Modbus address
Fieldbus1 parity
Unsigned char
RW
0x0FAC
4012
0x0FAD
4013
Modbus parity
Sensor differentiator dn
Float
RW
0x8158..0x8159
33112..33113
0x8159..0x815A
33113..33114
Sensor differentiator up
Float
RW
0x8160..0x8161
33120..33121
0x8161..0x8162
33121..33122
Capacity 100%
Float
RW
0x8168..0x8169
33128..33129
0x8169..0x816A
33129..33130
Fluid name
String (10 bytes)
RW
0x8188..0x818C
33160..33164
0x8189..0x818D
33161..33165
Capacity unit
String (7 bytes)
RW
0x81F8..0x81FB
33272..33275
0x81F9..0x81FC
33273..33276
Fmeasure
Float
R
0xA100..0xA101
41216..41217
0xA101..0xA102
41217..41218
FSetpoint
Float
RW
0xA118..0xA119
41240..41241
0xA119..0xA11A
41241..41242
Temperature
Float
R
0xA138..0xA139
41272..41273
0xA139..0xA13A
41273..41274
See addr 0x0427
Capacity 0%
Float
RW
0xA1B0..0xA1B1
41392..41393
0xA1B1..0xA1B2
41393..41394
Device type
String (6 bytes)
R
0xF108..0xF10A
61704..61706
0xF109..0xF10B
61705..61707
BHTModel number
String (14 bytes)
RW
0xF110..0xF116
61712..61718
0xF111..0xF117
61713..61719
Serial number
String (16 bytes)
RW
0xF118..0xF11F
61720..61727
0xF119..0xF120
61721..61728
Customer model
String (16 bytes)
RW
0xF120..0xF127
61728..61735
0xF121..0xF128
61729..61736
Firmware version
String (5 bytes)
R
0xF128..0xF12A
61736..61738
0xF129..0xF12B
61737..61739
Usertag
String (13 bytes)
RW
0xF130..0xF136
61744..61750
0xF131..0xF137
61745..61751
Valve output
Unsigned long
RW
0xF208..0xF209
61960..61961
0xF209..0xF20A
61961..61962
PID-Kp
Float
RW
0xF2A8..0xF2A9
62120..62121
0xF2A9..0xF2AA
62121..62122
PID-Ti
Float
RW
0xF2B0..0xF2B1
62128..62129
0xF2B1..0xF2B2
62129..62130
PID-Td
Float
RW
0xF2B8..0xF2B9
62136..62137
0xF2B9..0xF2BA
62137..62138
Density actual
Float
R
0xF478..0xF479
62584..62585
0xF479..0xF47A
62585..62586
Dynamic display factor
Float
RW
0xF508..0xF509
62728..62729
0xF509..0xF50A
62729..62730
Static display factor
Float
RW
0xF510..0xF511
62736..62737
0xF511..0xF512
62737..62738
Exponential smoothing
Float
RW
0xF520..0xF521
62752..62753
0xF521..0xF522
62753..62754
Fieldbus2 baudrate
Long integer
RW
0xFD48..0xFD49
64584..64585
0xFC49..0xFC4A
64585..64586
Modbus
baudrate OEMD2
Fieldbus1 baudrate
Long integer
RW
0xFD48..0xFD49
64840..64841
0xFD49..0xFD4A
64841..64842
Modbus
baudrate
Notes:
Access indicates whether parameter can be Read and/or Written.
When a byte parameter is read, the upper 8-bits of the Modbus register will be 0. When a byte
parameter is written, the upper 8-bits must be set to 0.
Long integer parameters have a length of 4 bytes and are mapped on two consecutive Modbus
registers. The first register contains bit 32-16, the second register contains bit 15-0.
Floating point parameters have a length of 4 bytes and are mapped on two consecutive Modbus
registers. Floats are in single precision IEEE format (1 sign bit, 8 bits exponent and 23 bits fraction).
The first register contains bit 32-16, the second register contains bit 15-0.
String parameters can have a length of maximal 16 bytes and can take up to 8 Modbus registers
where each register contains two characters (bytes). The upper byte of the first register contains the
first character of the string. When writing strings, the write action should always start from the first
register as a complete block (it is not possible to write a part of a string). If the string is shorter than
the specified maximum length the string should be terminated with a 0.
Details and meaning can be found in document 9.17.023 Operation instructions digital
instruments.
This document can be found at:
http://www.bronkhorst.com/en/downloads/instruction_manuals/