11. INFORMATION AND FAULT MESSAGES
11.1 Acoustic
signals
Beep sound
end of charging or discharging
Double beep
energy demand above or below energy control setting via RFS 2.1 / RFS 2.2 transmitter
11.2
Message in display
Fault 1
(F1)
No flash on release
→ Is the ignition wire correctly connected? Is the flash tube correctly installed?
Otherwise replace flash tube
Fault 2
(F2)
Afterglow
→ Switch the unit off and then on again. If this fault occurs repeatedly, the flash tube
is defective
Fault 3
(F3)
Electronic defect
→ Unit must be sent to Service for repair
Fault 4
(F4)
Over-voltage in flash capacitor
→ Unit must be sent to Service for repair
Fault 5
(th)
Thermal alarm, capacitors overheated
→ A cooling phase of approximately 20 minutes is started. Do not switch the unit off!
* Error messages are always also indicated by the red cognition light
12. FAQS
12.1
Changing flash tubes
1) Switch the unit off and remove the battery.
2) Wait for one minute for the flash capacitor to discharge. Remove the safety glass by pulling it
out of its retaining clips.
3) Press the contact spring downwards and pull the flash tube out carefully.
1)
2)
3)
Содержание Siros 400 L
Страница 2: ...Sandro B bler Switzerland...
Страница 4: ......
Страница 5: ......
Страница 9: ......
Страница 27: ......
Страница 28: ......
Страница 29: ......
Страница 33: ......
Страница 51: ......
Страница 52: ......
Страница 53: ......
Страница 57: ......
Страница 75: ...Thomas Lavelle France...
Страница 76: ...Bron Elektronik AG CH 4123 Allschwil Switzerland www bron ch BA112 00 Printed in Germany 10 16...