4. INSTALLATION
Locating the Cabinet
Ensure the following instructions are observed as this is vital for proper and trouble-free operation.
• Do not install the cabinet outdoors.
• Locate the appliance in a dry and well ventilated place.
• The unit is a commercial designed product, Do not install in noise sensitive areas.
• The unit must be positioned on a level surface.
• Do not install in the vicinity of heating radiators or other heat sources.
• Do not install the appliance where it is exposed to direct sunlight.
• Do not install the appliance where it is exposed to warm draughts or open doors.
• Keep the air intake and outlets free of obstructions.
• Adequate allowance should be made for door openings.
Ventilation
Ensure the following instructions are observed as this is vital for proper and trouble-free operation.
• Allow 150mm clearance between walls and the appliance.
• Ensure that the condenser fan is free to take in ambient air from the room.
• Adherence to minimum clearances is necessary
NOTE:
The maximum ambient temperature for the refrigerator is 30°C. Conditions
outside this temperature will result in cabinet temperatures outside the required 2
to 8°C range. Bromic recommends that the refrigerator is installed at an ambient
temperature of 20 to 25°C and relative humidity of 55%.