3. PRODUCT OVERVIEW

MED0220GD 

MED0374GD

Feature

Unit, Brand, Description

Data

Data

External Dimension

Height, mm

1310

1840

Width, mm

595

595

Depth, mm

598

598

Internal Dimension

Height, mm

1025

1555

Width, mm

505

505

Depth, mm

462

462

Capacity

Litres

215

372

Weight

kg

65

81

Air Circulation

Fan-Forced

Fan-Forced

Climate Class

4

4

Temperature

Range

+2°C to 8°C

+2°C to 8°C

Min/Max Temp Recorder

Yes

Yes

Display

Digital

Digital

Exterior 

Material

Powder coated sheet metal

Powder coated sheet metal

Electrical

Voltage, V

240

240

Frequency, Hz

50

50

Input Power, W

235

350

Current, A

2

2.7

Door

Solid or Glass

Low-E Glass + Argon Gas

Low-E Glass + Argon Gas

Shape

Flat

Flat

Self-Closing

Yes

Yes

Door Lock

Yes

Yes

Door Switch

Yes

Yes

Shelves

Qty

4

6

Type

Open rung shelf

Open rung shelf

Depth, mm

400

400

Adjustable

Yes

Yes

Capacity, kg

20

20

Internal LIghting

Internal Light

Yes

Yes

Alarms

Temperature

Yes

Yes

Audible

Yes

Yes

Compressor

Refrigerant

R134a

R134a

Charge, g

165

200

Defrost

Type

Off Cycle

Off Cycle

Manual/Automatic

Automatic

Automatic

Plug

Type

3-pin AU plug

3-pin AU plug

Rated Current, A

10

10

Feet/Wheels

Type

2 rollers at rear, 2 feet at front

2 rollers at rear, 2 feet at front

Содержание Medifridge MED0220GD

Страница 1: ...22 03 2022 V2 MED0220GD MEDO374GD ...

Страница 2: ...hout prior notification This Installation operation and service manual may not be reproduced in any form with prior written consent from Bromic Refrigeration 1 PREFIX 3 2 IMPORTANT NOTES WARNINGS 4 3 PRODUCT OVERVIEW 5 4 INSTALLATION 6 5 OPERATION 8 6 PRODUCT LOADING AND STORAGE 10 7 MAINTENANCE 11 8 TROUBLESHOOTING 12 9 ENVIRONMENTAL DISPOSAL NOTES 13 10 WARRANTY 14 ...

Страница 3: ...er strict control Refer to the National Vaccine Storage Guidelines Strive for 5 for more information on vaccine storage and handling or contact your local state of territory health department https www health gov au resources publications national vaccine storage guide lines strive for 5 MIXING OF VACCINE PRODUCTS WITH OTHER PRODUCT TYPES IS NOT RECOMMENDED The loading procedure of fridges contain...

Страница 4: ...ack of experience and knowledge unless they have been given supervision or instruction concerning use of the appliance by a person responsible of their safety Keep packaging material out reach of children Children should be carefully supervised to ensure that they do not play with the appliance Do not attempt to alter the appliance in any manner Please pay attention to the maximum load limit when ...

Страница 5: ...Powder coated sheet metal Electrical Voltage V 240 240 Frequency Hz 50 50 Input Power W 235 350 Current A 2 2 7 Door Solid or Glass Low E Glass Argon Gas Low E Glass Argon Gas Shape Flat Flat Self Closing Yes Yes Door Lock Yes Yes Door Switch Yes Yes Shelves Qty 4 6 Type Open rung shelf Open rung shelf Depth mm 400 400 Adjustable Yes Yes Capacity kg 20 20 Internal LIghting Internal Light Yes Yes A...

Страница 6: ...stall the appliance where it is exposed to warm draughts or open doors Keep the air intake and outlets free of obstructions Adequate allowance should be made for door openings Ventilation Ensure the following instructions are observed as this is vital for proper and trouble free operation Allow 150mm clearance between walls and the appliance Ensure that the condenser fan is free to take in ambient...

Страница 7: ...able is rated 70 C It is recommended that the cable should not come into contact with parts over 50 C Do not turn off power before consulting the person responsible Mark the power source clearly so the refrigerator is not unplugged or turned off accidentally Before Operating If the appliance has been sharply tilted or transported in the horizontal position it must be left to stand in the upright p...

Страница 8: ...gged in The electronic controller will display the current cabinet temperature after the unit has been connected to the power supply The LED indicates that refrigeration circuit is operating During the start up delay 3 minutes the LED is flashing Setting the temperature NOTE The controller is factory set at 4 C The controller is programmed to control cabinet temperatures between 2 to 8 C Controlle...

Страница 9: ...d the maximum and minimum temperatures two 2 times a day Recordings should be taken before the first door opening of the day MIN Temperature i Press the arrow key ii The Lo message will be displayed followed by the minimum temperature recorded iii Press the arrow key again or wait 5 seconds the normal display will be displayed MAX Temperature i Press the arrow key ii The Hi message will be display...

Страница 10: ... minimum Limit door opening durations to less than 30seconds Wait 10 minutes for product and cabinet temperature to stabilise before opening door again Ensure products and materials are between 2 and 8 C before loading if there is a breach in the cold chain isolate the affected products Do not use or discard until advice is given Some products are sensitive to ultraviolet UV light please see stora...

Страница 11: ...ve the products and transfer to a suitable location prior to cleaning the inside of the cabinet Do not use any aggressive or abrasive cleaning agents Never use a water hose or a high pressure jet to clean Do not use sharp or pointed objects for cleaning Clean the inside and outside of the appliance with a clean damp cloth or with a neutral cleanser Dry all surfaces with a dry cloth WARNING Disconn...

Страница 12: ...ate ventilation C Wall clearances below minimum distance D Dirty condenser coil E Ice build up on evaporator coil A Relocate unit B Keep ventilation slots free C Adjust to minimum clearance D Clean E Perform extended defrosting Cabinet Temperature Too Low A Set temperature too low A Adjust set temperature Cabinet lights not working A Light not switched on B Light is faulty A Switch on B Replace li...

Страница 13: ...ons must be observed In the interest of preventing environmental nuisance and promoting re use of the high quality materials employed used commercial refrigeration equipment must be collected for expert disposal and recycling Therefore observe the following disposal regulations The cabinet must not be discarded with regular waste The refrigeration circuit must not be damaged Before disposal make i...

Страница 14: ...e of any stock Failure to follow instructions may result in loss of or damage to stock Please refer to the National Vaccine Storage Guidelines Strive for 5 for more information on vaccine storage and handling or contact your local state or territory health department https www health gov au resources publications national vaccine storage guidelines strive for 5 BROMIC CANNOT BE RESPONSIBLE FOR THE...

Страница 15: ...NOTES ...