14
-
once sections are seen to be cut as required proceed to cut sections
NB:
For best results the wax cassettes should be cooled to their optimum sectioning
temperature
Keep the knife clear of debris by brushing upwards with the knife cleaning brush. NEVER
brush along the knife, always brush upwards.
3. GENERAL
MAINTENANCE
3.1 Daily
Care
Routine daily care consists of removing sectioning debris from the working area, brushing
debris from the knife and cleaning as appropriate.
-
first remove the microtome knife/blade holder and clean it (see section 3.4)
-
remove the debris trays and clean
-
wipe covers and remove excess debris as required
-
wipe visor if required
Note that the microtome requires no other routine maintenance.
3.2
Microtome Knives and Blades
Great care must be exercised when handling knives and or disposable blades.
-
knives/blades must be stored in their boxes when not in use
-
knives /blades fitted to the microtome must be properly guarded
-
particular care must be taken during cleaning and knife sharpening