4.3 Calibration Menu 

The scale is configured from the factory to match the specified settings for each unit, as defined by the 

product specifications and sales brochure. Modification of the settings can be accomplished by altering 

the calibration settings below. 

 

 

Calibration and/or configuration of calibration settings of your scale should be 

accomplished by a trained service technician using certified weights to ensure proper 

operation and accuracy. Calibration is not covered under warranty. 

 

4.3.1 Calibration Options 

In general weighing mode, press the 

TARE

 and 

ZERO/ON/OFF

 

keys. Release the keys when the 

“CAL” message pops up to open the Calibration Menu. The number of calibrations is shown. 

 

Navigate the menu using the          keys, press the 

TARE

 key to confirm or the 

ZERO/ON/OFF

 

key to cancel the operation / exit the menu. 

 

CAL 

SubMenu1 

SubMenu2 

Option 

Description 

ZERO 

 

 

Adjusts the CAL.P0 point only. 

LINE 

CAL.P0 

 

Stores the Zero calibration point (mandatory). 

CAL.P1 

 

Adds the cal. point  1  (mandatory). Use a 

standard weight over 10% F.S. 

END.Y 

YES / NO 

Saves and exits the calibration points:  

Yes 

= go to CAL.END;  

No 

= add another calibration point 

CAL.P2 

 

Adds the cal. point 2 (optional). Use a standard 

weight

 

over 10% F.S. and load used in 

CAL.P1

END.Y 

YES / NO

 

Saves and exits the calibration points:  

Yes 

= go to CAL.END;  

No 

= add another calibration point 

CAL.P3 

 

Linear calibration point3: do third weight point 

calibration, standard weight must be over 

10%FS and be larger than it in 

CAL.P2

, this 

point can be omitted. 

GEO 

 

16 

Do not use. 

See SBI 210-LCD Operation Manual. See 

Confi-

Geo.Cal

 

INPUT 

 

 

Do not use. 

See SBI 210-LCD Operation Manual. 

CAL.END 

 

 

Saves and exits the calibration points. The 

indicator reboots in weighing mode. 

Содержание 810036380263

Страница 1: ...Model MS 1000 LCD Operator Manual AWT35 100089 Rev AA...

Страница 2: ...roduct names used within this document are trademarks or registered trademarks of their respective holders Groupe de soci t s Avery Weigh Tronix 2021 Tous droits r serv s Aucune partie de cette public...

Страница 3: ...mple Weighing 9 3 2 Tare Weighing 9 3 3 Changing Units 9 3 4 BMI Function 10 3 5 Print Function Com Ports 10 3 6 Hold Function 10 3 6 1 Automatic Hold Mode Default 11 3 6 2 Average Hold Mode 11 3 6 3...

Страница 4: ...AWT35 100089 Rev AA 4 25 Manual revision history Current Issue Date Created Details of Changes AWT35 100089 Rev AA Sep 2021 New...

Страница 5: ...he mains lead must be connected to a supply outlet with a protective earth contact The electrical supply at the socket outlet must provide over current protection of an appropriate rating For your pro...

Страница 6: ...place the MS1000 platform in a location that will not degrade its accuracy Try to avoid placing the scale in direct sunlight or near air vents Place the scale on a level flat surface It is not advise...

Страница 7: ...of measure Battery status or AC DC operated 2 5 2 Operation Keys The key functions are listed below In menu mode the keys have secondary functions Enters exits the HOLD mode In Menu mode go to previou...

Страница 8: ...AWT35 100089 Rev AA 8 25 Zeroes the scale turns the scale on off In Menu mode cancel an operation and exit from the current working mode...

Страница 9: ...wer up the scale and zero the display if necessary by pressing the ZERO ON OFF key Be sure the scale is displaying weight in your preferred unit of measure Press the UNIT key if necessary 2 Place the...

Страница 10: ...full duplex RS 232 serial port and USB port designed for connection to either a PC or a serial printer The Print function sends the data out to the RS232 USB ports Use the correspondent adapter to con...

Страница 11: ...weighments 5 Press the HOLD key to exit the current mode If the weight oscillations exceed the HLD RNG division range value 5 div within the STB TIM time value 15 sec the indicator will display STB ER...

Страница 12: ...LD key to enable this function The Hold annunciator flashes 3 Step on the scale If the weight sensed is stable for STB TIM 15 s and above the NLD RNG zero range 10 div 5 lb 2 kg the indicator displays...

Страница 13: ...TARE key to confirm or the ZERO ON OFF key to cancel the operation exit the menu 4 1 1 User Options USER SubMenu1 SubMenu2 Option Default Description RESET NO NO NO Returns the USER parameter to the f...

Страница 14: ...GLE EH SCP SCP 12 Eh sp2 Lfuulf Lfuu Lf lf Lf uulf uu lf USB Port com2 BAUD RT 600 9600 Sets the baud rate of com2 1200 2400 4800 9600 19200 BYT FMT 8N1 8N1 Sets the bits format of com2 7O1 7E1 7O2 7E...

Страница 15: ...OMPAR mode NONE NONE COMPARE function modes NONE disabled See SBI 210 LCD Operation Manual and Config Compar L LOW IN LMT O HIGH OUT LMT HOLD Function HLD MOD NONE AUTO Enables the HOLD function mode...

Страница 16: ...div range can be averaged NLD RNG No Load Detected Range 1 255 10 Sets the weight range that the indicator considers as the Zero range for the relevant function e g Auto Hold Mode 1 255 the range of...

Страница 17: ...cator displays 2 TD 55 and the serial port continuously transmits 0x55 Use the and keys to change the character USB RD USB receive data testing same as RS232 receiving data testing USB TD USB send dat...

Страница 18: ...the TARE key to confirm or the ZERO ON OFF key to cancel the operation exit the menu CAL SubMenu1 SubMenu2 Option Description ZERO Adjusts the CAL P0 point only LINE CAL P0 Stores the Zero calibration...

Страница 19: ...platform 9 Press the TARE key to store the CAL P1 Point 10 Press the TARE key to save the calibration points and exit the Calibration mode or press the PRINT key to add CAL P2 3 points On calibration...

Страница 20: ...dary unit of measure 1 div 0 2 kg is automatically determined by the indicator according to the first unit PRIM Ut kg lb lb Selects the primary unit of measure kg or lb The default calibration unit is...

Страница 21: ...ght is outside the IZSM range DSP OVR if the initial zero is over range display the 0 error WEIGHT current weight CAL ZRO calibration zero LAST Z T last zero tare value WEIGHT CAL ZRO LAST Z T SAZSM 0...

Страница 22: ...25 COMPAR YES NO NO No disabled See SBI 210 LCD Operation Manual ACCUMU NO MANUAL AUTO NO No disabled See SBI 210 LCD Operation Manual GEO CAL YES YES Yes enables the Geographical Adjustment Factor i...

Страница 23: ...9V 600MA adaptor body with USA plugs included and 6xAA batteries not included Operating temperature range 32 104 F 0 40 C max relative humidity 80 at 86 F 30 C Communication interface RS 232 USB Port...

Страница 24: ...r 1 initial zero error 1 low battery 4 always 1 always 1 always 1 always 1 5 always 1 always 1 always 1 always 1 6 always 0 always 1 always 1 always 0 7 parity Parity parity Parity Command Response AS...

Страница 25: ...weighing range 1 Over capacity 2 0 Within weighing range 1 Under zero 3 0 Within range 1 Outside zero capture range 4 0 Not at center of zero 1 Center of zero 5 always 1 6 always 1 7 parity Command Re...

Страница 26: ...ty 1 Not at zero 0 Not over capacity 2 0 RAM error 1 ROM error 1 RAM okay 0 ROM okay 3 0 EEPROM error 1 Faulty calibration 1 EEPROM okay 0 Calibration okay 4 Always 1 Always 1 5 always 1 always 1 6 al...

Страница 27: ...D Simulate ZERO key no response from scale others Un known commands LF CR EH SP2 MM Status Byte gg Scale in motion GG Stable weight data Command Response ASCII HEX CR 0d Read scale weight P W1W2W3 dp...

Страница 28: ...ll free 800 637 0529 Tel 507 238 8702 Fax 507 238 8271 Email sales brecknellscales com http brecknellscales com Brecknell UK Foundry lane Smethwick West Midlands England B66 2LP Tel 44 0 8452 466717 F...

Отзывы: