Breathe Life 2000 Скачать руководство пользователя страница 16

Alarms and Troubleshooting

When an alarm notification occurs, there is a distinct sound and a display message corresponding to 

the priority level of the alarm. The priority level of an alarm is indicated by the color and the rate at 

which the message flashes. 

Please see next page for assistance in solving ventilator alarms.

N

OTE: 

 When attempting to resolve alarm conditions, ensure that the patient receives adequate 

ventilation therapy; place the patient on an alternate means of ventilation if necessary.

Ventilator Alarms

Medium-Priority Alarm

A yellow, steadily flashing alarm message is an alarm that indicates 

a potentially hazardous situation that must be resolved in a timely 

manner. Sound:  Sequence of three tones.

Low-Priority Alarm

A blue, non-flashing alarm message is an alarm that indicates a 

problem that is not hazardous but should be resolved.
Sound:  Single tone.

High-Priority Alarm

A red, rapidly flashing alarm message is an alarm that indicates a 

situation that requires immediate attention.
Sound:  Sequence of two sets of five tones.

How to Silence and Clear On-Screen Alarms

Alarm notifications that appear on the touch screen originate from the ventilator.
Silencing and cleaning on-screen alarms depends on alarm priority 
and number of active alarms.

         

Press the Silence Alarm button to temporarily silence the on-screen alarm for 

60 seconds. Pressing the Silence Alarm button silences only one alarm at a 

time. If more than one on-screen alarm occurs, press the Silence Alarm button 

once for each alarm. If the on-screen alarm is a medium or high priority alarm 

and is not silenced after 60 seconds, the alarm will continue with an additional 

buzzer.

A visible message will be displayed for High Priority alarms (High Temperature, 

High Circuit Pressure, and High PEEP alarms) When the alarm is resolved, 

touch Ok in message box to clear the message.

14

Содержание Life 2000

Страница 1: ...tructions for Use for a full list of warnings and cautions detailed operating instructions and additional information This guide is not a substitute for the Instructions for Use A copy of the Instruct...

Страница 2: ...Interface Life2000 Interfaces Breathe Universal Circuit connector System Setup Extended Range Configuration System Setup Stand Alone Configuration Starting Therapy Activity Buttons Home Screen Menu O...

Страница 3: ...ist of Tips Notes Cautions and Warnings The Breathe Technologies Instructions for Use can be downloaded at www breathetechnologies com healthcare providers resources or contact Breathe Technologies at...

Страница 4: ...ventilator 3 Belt clip for ventilator The belt clip is used to secure the ventilator when it is used in wearable configurations 4 Battery charger and AC power cord for the ventilator The battery charg...

Страница 5: ...tton 8 Battery charge indicator lights and battery charge scale 11 Handle 12 Air inlet and filter 13 Cooling filter cover 14 Cooling vents 15 Condensation tray 16 Power supply connection and cover 1 B...

Страница 6: ...ply and cord 1 Open the power supply connection cover and insert the AC power cord connector into the power supply connection on the back of the compressor until it clicks into place 2 Connect the ven...

Страница 7: ...met reposition the interface If problems persist try a different interface size Life2000 Interfaces The Breathe Pillows Entrainment Interface Checking the Breathe Pillows Interface Positioning 5 Wear...

Страница 8: ...rcuit connector 6 NOTE The interface assembly is packaged clean but not sterile The Breathe Universal Circuitconnector does not need to be cleaned or sterilized prior to first use Breathe interfaces a...

Страница 9: ...gas inlet connection on the ventilator when connected the quick connect end will click into place 5 Attach the COMBO2 oxygen tubing to the Breathe Pillows Entrainment interface Extended Range Configu...

Страница 10: ...utton Medium Activity Button Low Activity Button 8 1 Attach the oxygen regulator per the manufacturer s instructions for use onto the oxygen cylinder 2 Connect the oxygen hose to the Diameter Index Sa...

Страница 11: ...irm the selected Activity Icon is displayed at the bottom of the touch screen and the icon and volume are displayed at the top of the screen The ventilator will begin ventilating using the chosen pres...

Страница 12: ...r vibration 10 Time and date NOTE The Home Screen will initially display LPM when the ventilator is first powered on After an Activity Button is selected and ventilation begins the Home Screen will di...

Страница 13: ...mount of ventilator battery charge and the approximate time remaining NOTE The Ventilator Battery Charge Icon is always displayed on the touch screen during normal operation Ventilator Battery Charge...

Страница 14: ...essary check and empty the compressor s condensation tray and replace if necessary Every three to six months or more often if necessary check the compressor s air inlet filter and replace if necessary...

Страница 15: ...purge each port on the Interface Place the patient on an alternate means of ventilation if necessary With the Interface fully disconnected from the ventilator submerge the patient side of the Interfa...

Страница 16: ...ous but should be resolved Sound Single tone High Priority Alarm A red rapidly flashing alarm message is an alarm that indicates a situation that requires immediate attention Sound Sequence of two set...

Страница 17: ...Breath Rate Check patient s breathing rate Check for leakage or occlusion Adjust trigger sensitivity High Delivery Pressure Check the interface tubing for pinches and kinks Replace the interface if al...

Страница 18: ...operating Discontinue use of the compressor Place the patient on an alternate means of ventilation if necessary or connect the ventilator to an alternate pressure source oxygen cylinder or wall source...

Страница 19: ...art Guide Required for the Life2000 Ventilation System Patient Setup Setup By Company Name Company Phone Number Clinician Name Setup Date Patient Setup Checklist Life2000 Ventilator Life2000 Compresso...

Страница 20: ...A Released April 2019 2019 Breathe Technologies Inc Breathe Breathe Pillows Interface Breathe Technologies Life2000 and Universal Circuit are trademarks of Breathe Technologies Inc All Rights Reserve...

Отзывы: