Cleaning & Maintenance

Preventive Maintenance

Contact your Breathe Technologies® service representative to make arrangements for preventive maintenance, 

service, and component replacement per the chart below. The ventilation system can only be serviced or 

repaired by Breathe Technologies® or an authorized service center. 

Environmental Specifications

Do not use the ventilation system if the ambient temperature is greater than 40°C (104°F) or less than 5°C (41°F). 

Store the ventilation system in ambient temperatures less than 60°C (140°F) and greater than -20°C (-4°F).

  WARNING:

The rear housing of the ventilator enclosure may reach 49°C in a 40°C environment.

Cleaning 

Life2000® Ventilation System

Once a week (or more often if necessary): 

clean and disinfect the ventilation system using a clean 

cloth, clean and disinfect the external surfaces of the ventilation system with 70% isopropyl alcohol. 

Wipe the surface of the ventilation system with clean dry cloth to remove any residual cleaner.

Daily (or more often if necessary):

 

check and empty the compressor’s condensation tray and replace 

if necessary. 

Every three to six months (or more often if necessary):

 

check the compressor’s air inlet filter and 

replace if necessary.

Every three to six months (or more often if necessary): 

check the compressor’s cooling filter 

assembly and replace if necessary.

 

COMBO

2

®

 Hose

If dirt is visible on the outside of the hose, use a clean cloth and mild detergent such as dish-washing 

soap to remove it.

 

Replace every six months.

  WARNING:

Do not subject Breathe source gas supply hoses to heat sterilization, hot water pasteurization, 

autoclaving, radiation sterilization, ethylene oxide gas sterilization, or attempt to clean them in a 

dishwasher or microwave oven. Doing any of these may damage the hoses and impair gas delivery.

Breathe Pillows Entrainment Interface™

Breathe Technologies recommends a 

30-day replacement 

schedule for the Breathe Pillows 

Entrainment Interface. 

Once a week (or more often if necessary): 

Clean and purge the Breathe 

Pillows Entrainment Interface. 

12

REVISION

VENTILATOR 

COMPRESSOR

Before Revision C

2.5 Years from ship date

1 Year from ship date

Revision C and after

2.5 Years from ship date

2.5 Years from ship date

REV C

Содержание Life 2000

Страница 1: ...tructions for Use for a full list of warnings and cautions detailed operating instructions and additional information This guide is not a substitute for the Instructions for Use A copy of the Instruct...

Страница 2: ...Interface Life2000 Interfaces Breathe Universal Circuit connector System Setup Extended Range Configuration System Setup Stand Alone Configuration Starting Therapy Activity Buttons Home Screen Menu O...

Страница 3: ...ist of Tips Notes Cautions and Warnings The Breathe Technologies Instructions for Use can be downloaded at www breathetechnologies com healthcare providers resources or contact Breathe Technologies at...

Страница 4: ...ventilator 3 Belt clip for ventilator The belt clip is used to secure the ventilator when it is used in wearable configurations 4 Battery charger and AC power cord for the ventilator The battery charg...

Страница 5: ...tton 8 Battery charge indicator lights and battery charge scale 11 Handle 12 Air inlet and filter 13 Cooling filter cover 14 Cooling vents 15 Condensation tray 16 Power supply connection and cover 1 B...

Страница 6: ...ply and cord 1 Open the power supply connection cover and insert the AC power cord connector into the power supply connection on the back of the compressor until it clicks into place 2 Connect the ven...

Страница 7: ...met reposition the interface If problems persist try a different interface size Life2000 Interfaces The Breathe Pillows Entrainment Interface Checking the Breathe Pillows Interface Positioning 5 Wear...

Страница 8: ...rcuit connector 6 NOTE The interface assembly is packaged clean but not sterile The Breathe Universal Circuitconnector does not need to be cleaned or sterilized prior to first use Breathe interfaces a...

Страница 9: ...gas inlet connection on the ventilator when connected the quick connect end will click into place 5 Attach the COMBO2 oxygen tubing to the Breathe Pillows Entrainment interface Extended Range Configu...

Страница 10: ...utton Medium Activity Button Low Activity Button 8 1 Attach the oxygen regulator per the manufacturer s instructions for use onto the oxygen cylinder 2 Connect the oxygen hose to the Diameter Index Sa...

Страница 11: ...irm the selected Activity Icon is displayed at the bottom of the touch screen and the icon and volume are displayed at the top of the screen The ventilator will begin ventilating using the chosen pres...

Страница 12: ...r vibration 10 Time and date NOTE The Home Screen will initially display LPM when the ventilator is first powered on After an Activity Button is selected and ventilation begins the Home Screen will di...

Страница 13: ...mount of ventilator battery charge and the approximate time remaining NOTE The Ventilator Battery Charge Icon is always displayed on the touch screen during normal operation Ventilator Battery Charge...

Страница 14: ...essary check and empty the compressor s condensation tray and replace if necessary Every three to six months or more often if necessary check the compressor s air inlet filter and replace if necessary...

Страница 15: ...purge each port on the Interface Place the patient on an alternate means of ventilation if necessary With the Interface fully disconnected from the ventilator submerge the patient side of the Interfa...

Страница 16: ...ous but should be resolved Sound Single tone High Priority Alarm A red rapidly flashing alarm message is an alarm that indicates a situation that requires immediate attention Sound Sequence of two set...

Страница 17: ...Breath Rate Check patient s breathing rate Check for leakage or occlusion Adjust trigger sensitivity High Delivery Pressure Check the interface tubing for pinches and kinks Replace the interface if al...

Страница 18: ...operating Discontinue use of the compressor Place the patient on an alternate means of ventilation if necessary or connect the ventilator to an alternate pressure source oxygen cylinder or wall source...

Страница 19: ...art Guide Required for the Life2000 Ventilation System Patient Setup Setup By Company Name Company Phone Number Clinician Name Setup Date Patient Setup Checklist Life2000 Ventilator Life2000 Compresso...

Страница 20: ...A Released April 2019 2019 Breathe Technologies Inc Breathe Breathe Pillows Interface Breathe Technologies Life2000 and Universal Circuit are trademarks of Breathe Technologies Inc All Rights Reserve...

Отзывы: