Using Oxygen

1. Mount the oxygen tube to the 

connector.

2. Connect the oxygen tube. A click 

sound is heard when the tube is 

properly attached.

Disconnect:

1. Press the release button to  

disconnect the oxygen tube.

2. Remove the tube.
3. Turn off the oxygen source.

Replacing the Patient Air Inlet 

Filters

The filters are located in the filter 

cassette at the side of the ventilator. 

 

Washable Filter

 (grey): Replace at 

least once a year, wash at least once 

a week.

 

Disposable Filter

 (white): Replace 

at least every 4th week, or more 

often in high pollution or pollen-rich 

environments.

Replacing the Cooling Air Filter

The cooling air inlet filter is located 

in the filter cassette at the side of 

the ventilator. Replace the filter at 

least once a year. Wash the filter at 

least once a week.

• O

2

 must be turned off when the Vivo 65 

is not in operating mode!

• It is recommended to monitor the  

oxygen concentrations.

• Oxygen flow must not exceed 15 l/min.

Symbols Used in the Menu

SYMBOL DESCRIPTION

Internal battery level

Click-on battery level

Home Mode activated

Dual limb circuit for integrated exhalation 

valve selected (Dual)

Single limb circuit with leakage port  

selected (Leakage)

Single limb circuit with active exhalation 

valve selected (Exh. valve)

MPV breath mode selected. Circuit with 

mouthpiece interface shall be used.

SpO

2

 connected

FiO

2

 connected

CO

2

 connected

Multiple pages

Multiple content available

High priority alarm event in history list

Medium priority alarm event in history list

Företagsvägen 1 · SE-435 33 Mölnlycke · Sweden

Phone +46 31 86 88 00

www.breas.com

*Vivo by Breas, Breas and the Breas logo are trademarks of Breas Medical AB. © 2018 Breas Medical – All rights reserved. Breas Medical reserves the right to make changes in specifications and features shown herein, or discontinue the product described 
at any time without notice or obligation. Contact your Breas representative for the most current information. MAR-000265.v.0.0

Human Design Medical
1-844-848-6872

Connecting and  

Disconnecting Cables

Insert the connector with 

the marking pointing  

upwards.

Pull the connector sleeve, 

not the cable itself or 

cable restrainer to release 

the connector.