3

Replacing the Patient Circuit

Stop treatment before disconnecting the 
patient circuit.

Connect a Single Limb Circuit with Active  

Exhalation Valve:

1. Connect the patient circuit to the patient air outlet 

of the ventilator.

2. Connect the small tubing to the exhalation valve 

control pressure outlet of the Vivo 65.

3. Connect the other end of the patient circuit to an 

HME or patient interface.

Connect Dual Limb 

Circuit 

1. Connect one patient 

tube to the patient air 

outlet of the ventilator.

2. Connect the other  

patient tube to the 

dual limb circuit.

3. Connect the other end 

of the patient circuit 

to an HME or patient 

interface.

Attach the Click-on Battery

1. Use a thin stick to 

open the cover for  

the click-on battery 

connectors. 

2. Make sure the cover is  

completely opened.

3. Hold the Vivo 65 as 

shown in the picture, 

in order to target 

the click-on battery 

holders (marked with 

circles).

4. Tilt the Vivo 65 into an 

upright position.

5. Once in position, a 

clicking sound should 

be heard.

6. Use the screw to 

secure the click-on 

battery by pressing it 

in and turn clockwise.