2

Connect Patient Circuit Insert

The choice of insert makes it possible to connect the 

Vivo 65 to either a dual limb pediatric circuit, a dual 

limb adult circuit or a single limb circuit. The Vivo 65 

automatically detects which insert has been installed. 

Before connecting an insert, make sure that the  

configurations of patient mode and patient circuit type 

are correct:

See users manual for complete instruc-

tions.

• To optimize the treatment of the Vivo 65’s  

pediatric settings in Pediatric mode when 

using a dual limb circuit we recommend to 

use a pediatric dual limb insert and a 15 mm 

dual limb circuit. 

2. Lock the insert by 

turning the lock-ring 

clockwise, so that the 

indicator points to the 

lock symbol on the 

ventilator.

3. Make sure a “click” 

sound is heard when 

fastening the lock-ring. 

This sound ensures that 

the insert is correctly 

attached.

Unlock and disconnect in 

reverse order.

• Always perform a 

pre-use test after 

changing the pa-

tient circuit type 

and/or insert.

• Do not remove 

the gasket that 

is placed inside 

of the Vivo 65’s 

insert socket. 

Företagsvägen 1 · SE-435 33 Mölnlycke · Sweden

Phone +46 31 86 88 00

www.breas.com

Human Design Medical
1-844-848-6872

INSERT TYPE 

PATIENT MODE 

PATIENT CIRCUIT TYPE

Dual limb insert 
Adult

Adult

• Dual limb circuit with 

integrated exhalation 
valve (Dual)

• Single limb circuit with 

leakage port (Leakage)

• Circuit with mouthpiece 

interface

Pediatric 

Dual limb insert 
Pediatric

• Dual limb circuit with 

integrated exhalation 
valve (Dual)

• Single limb circuit with 

leakage port (Leakage)

• Circuit with mouth-

piece interface

Single limb insert
(optional)

Adult / Pediatric 

• Single limb circuit with 

active exhalation valve 
(Exh. valve)

• Single limb circuit with 

leakage port (Leakage)

• Circuit with mouthpiece 

interface

1. Connect the insert to 

the ventilator. While 

connecting, the lock 

indicator on the insert’s  

lock-ring should point 

to the unlock symbol 

on the Vivo 65.

Removing the gasket will result in a patient 

circuit leakage and a Disconnection alarm and/or 

Exhalation Valve Control Error alarm.