Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
4–3
PV 101+/PV 102+ Service Manual
CALIBRATION INSTRUCTIONS
4.4
Verifying the pressure calibration
1
Start the PV 101+/PV 102+.
2
Remove the casing (see section 3.2).
3
Connect a pressure gauge to the patient air inlet and arrange a leakage with a
5 mm flow restrictor (see diagram below).
4
Set both the IPAP and EPAP pressure to 20 mbar.
5
Check that the pressure rises to 20 mbar ±1 mbar. If not, the P204 potentiometer
must be adjusted.
6
Set both the IPAP and EPAP pressure to 4 mbar.
7
Check that the pressure rises to 4 mbar ±0.5 mbar. If not, the P203 potentiometer
must be re-adjusted.
8
If necessary, repeat the pressure calibration procedure (see section 4.3).
9
Remove the equipment and reassemble the casing (see section 3.2).
mbar
Pressure gauge
5 mm leakage
Patient circuit
Plug