MAINTENANCE SERVICE INSTRUCTIONS
PV 101+/PV 102+ Service Manual
2–8
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
Rate
3
Adjust the settings to:
IPAP
10 mbar
EPAP
4 mbar
Back-up rate 8 BPM (PV 102+ only)
The PV 101+ back-up rate is fixed to 6 BPM.
4
Measure the rate and check that it is correct.
2.6.5
Replacing the filters
The patient air filters are located in the filter holder under the carrying handle of the
device. The internal cooling filter is located in the lower casing.
1
Replace the washable grey air filter.
2
Replace the disposable white filter, if used.
3
Replace the internal cooling filter from underneath the machine.
2.6.6
Checking the accessories
1
Inspect the patient circuit. Replace it if necessary.
2
Inspect any other accessories.
2.6.7
Adjusting the settings for the patient
•
Adjust all the patient settings to the prescribed values.
mbar
Pressure gauge
5 mm leakage
Patient circuit
Plug