3
2
1
13
Remove cables 1 and 2. Then
disconnect cable 3 from the heated
wire board.
Twist when
Reassembling
14
Disconnect the tubes between the
FiO
2
sensor and the blower
assembly. Disconnect at the blower.
Note:
When reassembling, twist
the return tube about one turn
clockwise. This is for creating a
bow on the tube that prevents it
from being squeezed when
assembling the front cover.
3
4
15
Vivo 45 LS and Nippy 4+ only:
Disconnect the exhalation valve
tube (4).
NOTE
When assembling, the plastic tip (3)
is to be inserted into the hole at the
back of the blower. Ensure that its
seal doesn
’
t slide into the hole. Use
a torch to check.
Opening the Ventilator and Replacing the Main Components 41
Doc. 007132 D-2
Vivo 45/Nippy 4 Series Service manual
Содержание Nippy 4 Series
Страница 1: ...D 2 007132 En Service Manual...
Страница 2: ...DOC 007132 EN EU D 2 2020 12 22 Service manual Vivo 45 Nippy 4 Series...
Страница 3: ......
Страница 7: ......
Страница 26: ...3 6 View F 42 1 42 2 42 3 42 9 Parts Location 25 Doc 007132 D 2 Vivo 45 Nippy 4 Series Service manual...
Страница 27: ...3 7 Screw Kits 3 7 1 MRT M3x6 FZB 26 Parts Location Vivo 45 Nippy 4 Series Service manual Doc 007132 D 2...
Страница 96: ...7 321820 071326 Service Manual Vivo 45 Nippy 4...