2.7
Power Failure Test
2.7.1
Checking the Power Failure Alarm
NOTE
The
“
Low Alarm Battery
”
alarm shall not be active prior to the test.
1
Remove the internal battery as in 5.1
2
Connect the ventilator to mains.
3
Turn the ventilator on and start treatment.
4
Disconnect mains power supply.
5
Make sure that the ventilator gives the power failure alarm. The power failure alarm
is given with an audible signal and the red alarm LED flashing. The alarm shall be
given for, at least, two minutes. If the power fail alarm stops within two minutes,
change the super capacitors (see section 5.15
6
Reconnect the internal battery as in 5.1
2.8
Complete Function Test
To perform the complete function test you need the Service Software, which can be
downloaded from the Breas Extranet. Also, you will need the equipment listed in the
section on tools above.
Contact Breas technical support for more information.
2.9
CO
2
Sensor Gas Span Check
If a CO
2
sensor is used with the ventilator it shall be checked once a year to verify the
sensor readings.
To perform the CO
2
gas span check you need CO
2
calibration gas consisting of:
•
5% CO
2
(carbon dioxide)
•
21% O
2
(oxygen)
•
74% N
2
(nitrogen)
You will also need a gas regulator.
Vivo 45/Nippy 4 Series Service manual
Doc. 007132 D-2
Содержание Nippy 4 Series
Страница 1: ...D 2 007132 En Service Manual...
Страница 2: ...DOC 007132 EN EU D 2 2020 12 22 Service manual Vivo 45 Nippy 4 Series...
Страница 3: ......
Страница 7: ......
Страница 26: ...3 6 View F 42 1 42 2 42 3 42 9 Parts Location 25 Doc 007132 D 2 Vivo 45 Nippy 4 Series Service manual...
Страница 27: ...3 7 Screw Kits 3 7 1 MRT M3x6 FZB 26 Parts Location Vivo 45 Nippy 4 Series Service manual Doc 007132 D 2...
Страница 96: ...7 321820 071326 Service Manual Vivo 45 Nippy 4...