2
Connect the patient circuit.
3
Switch on the ventilator and make sure it operates normally.
2.6
Internal Inspection
WARNING!
Make sure to disconnect the power supply before opening the casing of the ventilator.
CAUTION!
Always perform an internal inspection after having opened the ventilator.
2.6.1
Cleaning the Inside of the Ventilator
1
Open the casing. See the section on opening the ventilator for instructions.
2
Remove any dirt or dust that has accumulated in the ventilator.
2.6.2
Checking the Cables
Inspect all the cables and their connectors. Make sure that the cables and the wires are
not pinched or kinked.
2.6.3
Checking the Fastening of Components
1
Make sure that all the components, such as the pneumatic block, the circuit boards
and the connectors are securely fastened.
2
Check that the blower assembly is placed correctly and that the vibration
dampening pad is properly in place.
3
Check that the cooling fan is placed correctly.
4
Make sure that all tubes are properly fastened and not pinched or kinked. Do not
forget the air inlet and outlet of the blower assembly.
2.6.4
Checking the Power Adapter
1
Check that the power adapter is undamaged.
2.6.5
Reassembling the Casing
See the section on opening the ventilator and replacing parts for instructions.
Doc. 007132 D-2
Vivo 45/Nippy 4 Series Service manual
Содержание Nippy 4 Series
Страница 1: ...D 2 007132 En Service Manual...
Страница 2: ...DOC 007132 EN EU D 2 2020 12 22 Service manual Vivo 45 Nippy 4 Series...
Страница 3: ......
Страница 7: ......
Страница 26: ...3 6 View F 42 1 42 2 42 3 42 9 Parts Location 25 Doc 007132 D 2 Vivo 45 Nippy 4 Series Service manual...
Страница 27: ...3 7 Screw Kits 3 7 1 MRT M3x6 FZB 26 Parts Location Vivo 45 Nippy 4 Series Service manual Doc 007132 D 2...
Страница 96: ...7 321820 071326 Service Manual Vivo 45 Nippy 4...