Trouble Shooting Tips
Battery is not charging
1.
Disconnect the
Z2
from the
Battery
and the power
supply.
2.
If you squeeze the arrows on the sides of the
PowerShell
, you should feel the
Z2
release.
3.
Reassemble the
Z2
and the battery to the
PowerShell
and to the power supply. When reassembling, make
sure you hear the components click together and
see that the DC plug (silver cylinder in the bottom of
PowerShell
) goes into the hole in the
Z2
.
4.
Make sure to fully charge the
Battery
for 12 hours
after use.
5.
The battery icon on the
Z2
screen will stop flashing
and all bars will be solid.
NOTE:
Battery icon starts showing 4 bars at 90% charge.
Wait for battery icon on
Z2
LCD screen to stop flashing.
Number of Solid Bars
Charge Remaining
1 10-35%
2
35-63%
3
63-90%
4
90-100%
Typically it will take 12 hours to fully charge the battery. After a few
minutes, the back light of the display will turn off. This is to con-
serve power. It does not indicate that the battery is fully charged.
NOTE:
The
PowerShell
will not charge the battery at the
same time as you are using your
Z2
for treatment.