9
What to do if …..
Problem
Reason
Solution
Heart rate
symbol
Appears in the measurement
condition and flashes when
pulse is detected.
• Measurement in progress, remain quiet.
Low battery
indicator
Appears when the battery
voltage is excessively low or the
positions of the batteries are
incorrect.
• Replace all four batteries with new ones.
Insert the batteries in the correct
positions. Be aware of +/- positions.
Measuring
error
Appears when the accurate
blood pressure and pulse could
not be obtained.
• Press “start/stop” button again and
remeasure.
• Check if cuff is wrapped according
to instructions.
• Check that there are no kinks in tube.
• Check that hand is relaxed.
• Check if talking or moving
during measurement.
• Check if posture is correct.
E1 shows
The cuff is not secure.
• Refasten the cuff and then measure again.
E2 shows
The cuff is very tight.
• Refasten the cuff and then measure again.
E 10 or
E 11 shows
The monitor detected motion,
talking or the pulse is too poor
while measuring.
• Relax for a minimum of 1 minute and then
measure again.
E 20 shows
The measurement process does
not detect the pulse signal.
• Ensure there is no clothing between the cuff
and the arm, and no constriction on the arm
from clothing, then measure again.
E 21 shows
Measure incorrectly.
• Relax for a minimum of 1 minute and then
measure again.
EE 3 - EE15
Error while measuring.
• Retake the measurement. If the problem
persists, contact the retailer or our customer
service department for further assistance.
Refer to the warranty for contact information
and return instructions.
English