Guidance and manufacturer’s declaration – electromagnetic emissions
The ME equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME equipment should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The ME equipment uses RF energy only for it’s internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class BF
Complies
Harmonic emissions
IEC 61000-3-2
Not Applicable
The ME equipment is solely battery powered.
Voltage fluctuations /
flicker emissions
Not Applicable
Non-Life Support Equipment Separation Distance Calculation
(3Vrms / 3V/m compliance)
Rated maximum output
power of transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
in ISM bands
80 MHz to
800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
Guidance and manufacturer’s declaration – electromagnetic immunity
The ME is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME should assure that it is used in such an environment.
Immunity test
IEC60601 test
level
Compliance level
Electromagnetic environment –
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
Complies
Floors should be wood,concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Radiated RF
IEC 61000-4-3
3 V/m 80MHz to
2.5GHz
Complies
Field strengths outside the shielded
location from fixed RF transmitters, as
determined by an electromagnetic site
survey, should be less than 3 V/m.
Interference may occur in the vicinity
of equipment marked with the follow-
ing symbol:
Separation distance calculation
provided above. If a known transmitter
is present the specific distance can be
calculated using the equations.
Conducted RF
IEC 61000-4-6
3Vrms 150kHz to
80MHz
Not Applicable
(no electrical
cabling)
Electrical fast transient
IEC 61000-4-4
±2kV power line
±1kV I/O lines
Not Applicable
The ME equipment is solely battery
powered.
Surge
IEC 61000-4-5
±1kV differential
±2kV common
Not Applicable
Power frequency magnetic
field
IEC 61000-4-8
3 A/m
Complies
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
Voltage dips, short
interrupts and voltage
variations on power
supply input lines
IEC 61000-4-11
>95% dip 0.5
cycle
60% dip 5 cycles
70% dip 25
cycles
95% dip 5 sec.
Not Applicable
The ME equipment is solely battery
powered.
d=[ ]
V
1
3.5
P
d=[ ]
E
1
3.5
P
d=[ ]
E
1
7
P
d=[ ]
V
1
3.5
P
d=[ ]
E
1
3.5
P
d=[ ]
E
1
7
P
d=[ ]
V
1
3.5
P
d=[ ]
E
1
3.5
P
d=[ ]
E
1
7
P