6
Aesculap
®
S4 MIS instruments
General information
Dried or affixed surgical residues can make cleaning
more difficult or ineffective and lead to corrosion.
Therefore the time interval between application and
processing should not exceed 6 h; also, neither fixating
pre-cleaning temperatures >45 °C nor fixating disin-
fecting agents (active ingredient: aldehydes/alcohols)
should be used.
Excessive measures of neutralizing agents or basic
cleaners may result in a chemical attack and/or to fad-
ing and the laser marking becoming unreadable visu-
ally or by machine for stainless steel.
Residues containing chlorine or chlorides e.g. in surgi-
cal residues, medicines, saline solutions and in the ser-
vice water used for cleaning, disinfection and steriliza-
tion will cause corrosion damage (pitting, stress
corrosion) and result in the destruction of stainless
steel products. These must be removed by rinsing thor-
oughly with demineralized water and then drying.
Additional drying, if necessary.
Only process chemicals that have been tested and
approved (e.g. VAH or FDA approval or CE mark) and
which are compatible with the product’s materials
according to the chemical manufacturers’ recommen-
dations may be used for processing the product. All the
chemical manufacturer's application specifications
must be strictly observed. Failure to do so can result in
the following problems:
■
Optical changes of materials, e.g. fading or discol-
oration of titanium or aluminum. For aluminum,
the application/process solution only needs to be
of pH >8 to cause visible surface changes.
■
Material damage such as corrosion, cracks, fractur-
ing, premature aging or swelling.
►
Do not use metal cleaning brushes or other abra-
sives that would damage the product surfaces and
could cause corrosion.
►
Further detailed advice on hygienically safe and
material-/value-preserving reprocessing can be
found at www.a-k-i.org, link to Publications, Red
Brochure – Proper maintenance of instruments.
Disassembling the product before carry-
ing out the reprocessing procedure
►
Disassemble the product immediately after use, as
described in the respective instructions for use.
►
Open up products with hinges.
Preparations at the place of use
►
If applicable, rinse non-visible surfaces preferably
with deionized water, with a disposable syringe for
example.
►
Remove any visible surgical residues to the extent
possible with a damp, lint-free cloth.
►
Transport the dry product in a sealed waste con-
tainer for cleaning and disinfection within 6 hours.
Preparation before cleaning
►
Dismantle the product prior to cleaning, see Disas-
sembling.
Содержание Aesculap S4 MIS Spyder
Страница 2: ...30 40 50 60 70 30 40 50 60 70 80 80 100 80 70 60 50 40 90 1 2 4 3 5 6 7 8...
Страница 3: ...FW734R 4 7 b FW238R 5 a 6 A B C D...
Страница 5: ...6 7 FW238R J K...
Страница 119: ...115 S4 TA012384 S4 Spinal System S4 FRI S4 SRI S4 MIS S4 S4...
Страница 122: ...118 Aesculap S4 MIS Aesculap https extranet bbraun com Aesculap 6 45 C...
Страница 123: ...119 VAH FDA CE pH 8 www a k i org Rote Brosch re 6...
Страница 124: ...120 Aesculap S4 MIS 55 C...
Страница 125: ...121 FW705R FW709R FW711R FW712R FW714R FW716R FW734R 20 FW712R...
Страница 126: ...122 Aesculap S4 MIS FW705R FW709R FW711R FW714R FW716R FW734R 20 FW705R FW709R FW714R FW716R...
Страница 127: ...123 BBraun Stabimed T C F t I 15 2 pH 9 II 1 III 15 2 pH 9 IV 1 V...
Страница 128: ...124 Aesculap S4 MIS I 15 1 5 II III 5 IV 5 V...
Страница 130: ...126 Aesculap S4 MIS FDA CE BBraun Stabimed I 15 1 5 II T C F t I 15 2 pH 9 II 1...
Страница 131: ...127 BBraun Helimatic Cleaner alcaline T C F t I 25 77 3 II 55 131 10 pH 13 5 P 0 5 pH 11 III 10 50 1 IV 90 194 5 V...
Страница 132: ...128 Aesculap S4 MIS Aesculap Aesculap STERILIT I JG600 STERILIT I JG598...
Страница 191: ...187 S4 TA012384 S4 S4 FRI S4 SRI S4 MIS K S4 S4 X FW271R FW258R...
Страница 194: ...190 Aesculap S4 MIS OP 6 45 C OP VAH FDA CE pH 8 www a k i org OP 6...
Страница 195: ...191 55...
Страница 198: ...194 Aesculap S4 MIS 1 15 1 5 II III 5 IV 5 V...