BpTRU Operator’s Manual
52
CAUTION
Stop using the monitor if there is evidence of any of
the damage listed below. Replacement of damaged
parts or accessories is mandatory for correct
functioning of the monitor.
Inspection
All labels on the BpTRU should be examined for
integrity and legibility (i.e., they should not be
peeling). Damaged or missing labels may lead to
misuse of the monitor.
The BpTRU and power supply with cord should be
examined for possible electrical hazards caused by
any loose connectors, cracks or other broken parts,
including damage which might permit access to the
interior of the chassis or power supply.
All electrical cables should be examined for possible
electrical hazards caused by cracked or broken
connectors or wires.
The BpTRU and power supply should be examined to
ensure that they do not rattle (indicating loose
components inside which may lead to faulty
operation).
Prior to using the monitor or after replacing the
battery, check for battery irregularities such as cracks,
deformation, leakage, corrosion, or other problems.
Discontinue use and replace the battery if any such
problems arise.
The coiled tubing, blood pressure cuffs, and their
connections should be examined for cracks or breaks
which may lead to air leakage during inflation and
deflation.
The blood pressure cuffs should be examined for
tears in the material and seams of the cuff which may
lead to the cuff rupturing when inflated. The cuffs
should also be inspected for legibility of cuff
Содержание BPM100
Страница 2: ......
Страница 14: ...Manuale di istruzioni BpTRU 8...
Страница 20: ...BpTRU Operator s Manual 14...
Страница 30: ...BpTRU Operator s Manual 24...
Страница 48: ...BpTRU Operator s Manual 42...
Страница 62: ...BpTRU Operator s Manual 56...
Страница 82: ...BpTRU Operator s Manual 76...
Страница 90: ...BpTRU Operator s Manual 84...