Display Reference
37
E10
Cuff over 300 mmHg
for more than 5
seconds.
Ensure patient is sitting quietly to
reduce motion artifacts during
inflation. Check cuff and tubing for
kinks.
E11
Cuff pressurized for
more than 130 seconds.
Check cuff and tubing for leaks due
to cracks and tears. Check connectors
for tightness. Check cuff and tubing
for kinks.
E12
Automatic cycling
timing violation.
Check cuff and tubing for kinks.
E13
Device reset.
Readings in memory, if any, are
available for review. Check monitor
environment for electromagnetic
interference (ESD, power
interruptions). Connect USB cable
between measurements or when
monitor is powered off.
E14
Data corrupted after
reset.
Power the BpTRU off, wait 5 seconds
and power on again.
E15
Low battery
Charge the battery
a.
In Automatic mode these errors are recorded in memory and mea-
surements continue. The measurement on which the error was
detected is not included in the average of measurements displayed at
the end of the series of measurements.
BP Analysis Errors (E20 to E27)
If an analysis error occurs in the determination of any BP
reading, an error code is displayed in the corresponding
display. All other valid readings are still displayed in the
corresponding displays.
If an analysis error occurs during Automatic mode, the
error code is stored in memory and the series
automatically continues until six measurements are
taken.
Содержание BPM100
Страница 2: ......
Страница 14: ...Manuale di istruzioni BpTRU 8...
Страница 20: ...BpTRU Operator s Manual 14...
Страница 30: ...BpTRU Operator s Manual 24...
Страница 48: ...BpTRU Operator s Manual 42...
Страница 62: ...BpTRU Operator s Manual 56...
Страница 82: ...BpTRU Operator s Manual 76...
Страница 90: ...BpTRU Operator s Manual 84...