sound after turning it on, please examine it and take some methods by following steps:
1. To examine the power wire is connected correctly and fully or not.
2. To find the position of “220V” (Figure 5) in the back of machine, find the 2 fuses(Figure 6),
and examine the using fuse. If the using fuse is broken, please replace it by the spare fuse.
Spare
fuse
Fuse
being
used
5
6
3. To fix the support bracket:
please fix the support bracket stably by screws, then put the
handle on it well.
4. Treatment Preparedness:
please turn on the machine, 2-3 minutes later, do the
treatment after the handle is cooling sufficiently.
Notice:
If press the button of handle to output the light immediately after turning on the machine,
the machine will be damaged and its life time will be shorter.
Please check the handle to
confirm if the sapphire’s position is frosting/ water coagulated, the handle is totally cooling;
if not, the handle is not cooling and freezing, please do not operate the machine , otherwise the
handle will become overheating and damaged. ( if the ipl handle is not cool and you can’t hear
the voice of water flow inside the handle, please contact us)
5. the reasons of no refrigeration
: (please put off the machine when check its system)
A. To check the water is added to machine or not.