on different patients’ skin situations.
9.3. OPT parameters adjustment
(1).
Switching treatment records
Treatment
records
are
used
to
display
the
current
treatment
parameters,
record different treatment parameters, easy to treat different patients, just click to change the
treatment records, there are three options; orange indicates the current selected treatment
records.
(
2).
Modify Treatment parameters
After clicking to modify treatment parameters, the original treatment mode display area will show
the up and down arrow to modify treatment parameters. After modifying treatment parameters,
press the number button to finish the modification.
The number of pulses can be adjusted by clicking directly, adjustment range 1-6;Pulse interval
can be adjusted by up and down arrow, range 1-4 seconds.
(3) Adjustment of RF working time
Click the radio button after working hours and the number of the original photon
pulses pulse delay button will turn
up and down arrows.
RF working time
adjustment
range:50-2500ms, through the up and down arrows to adjust its size, continue
pressing the adjustment button if the RF operating time will automatically increase or decrease,
RF step is to adjust working time: 50 ms, the larger the number, RF longer duration of
action, it is recommended to use between
300-900ms, can be adjusted according to actual situation. After the adjustment is complete click
radio button again working hours, up and down arrows will be reduced to the number of pulses
and pulse delay button.
(4). Adjustment of photon energy output
Photon energy output adjustment range 1-50 J/cm2, adjust its size by up and down
arrows, if regulation has been pressing the button, the photon energy will