7 Preparation
46
NightKNIFE / Ligator
MN031-447-S0-20101217-EN
HF cable
1. Check the connector for damage and corrosion.
2. Visually inspect the insulation for damage.
7.6.2. LIGATOR
inspection
DANGER
Risk of patient burns with brittle or defective insulation!
X
Replace instrument components with damaged
insulation.
Jaw insert
1. Check the jaws
A1
for proper opening and closing.
2. Visually inspect the ball on drawrod
A4
for damage.
3. Check the insulation of drawrod
A4
for damage.
4. Check whether the shank tube
A4
is bent.
Shank tube
1. Visually inspect the insulation for damage.
2. Check whether the shank tube
B
is bent.
Handle
1. Check the ratchet lever
C1
, locking lever
C3
and
buttons
C4
for ease of motion.
2. Check cable connector
C2
for damage and corrosion.
HF cable
1. Check the connector for damage and corrosion.
2. Visually inspect the insulation for damage.
MN031-447-S1 EN
2
X
HF cable
1. Check the connector for damage and corrosion.
2. Visually inspect the insulation for damage.