5 Operation
26
NightKNIFE / Ligator
MN031-447-S0-20101217-EN
Changing the output power of the HF device
X
Before increasing the output power of the HF device,
check that:
x
all HF cables and connectors are properly
connected;
x
the ligation instrument is properly connected
(see the operating instructions of the HF
device);
x
the foot switch works properly;
x
the insulation of the HF cable
D
, the jaws
A1
, the shank tube
B
and the trocar sleeve
is in good condition;
x
the active electrodes of the jaws
A1
are
clean and not worn out.
5.3. Withdrawal
WARNING
Risk of patient injury due to damaged or broken-off
parts!
X
Check the jaw insert after each withdrawal. All parts
must be present.
1. Close the jaws
A1
.
2. Withdraw the ligation instrument from the trocar sleeve.
MN031-447-S1 EN
the insulation of the HF cable
C3
, the jaws