7 Preparation
MN031-447-S0-20101217-EN
NightKNIFE / Ligator
45
7.6.1. NightKNIFE
inspection
DANGER
Risk of patient burns with brittle or defective insulation!
X
Replace instrument components with damaged
insulation.
Jaw insert
1. For NightKNIFE with exchangeable blade: install a new
blade (see section 4.1, page 18).
2. Check the jaws
A1
for proper opening and closing.
3. Visually inspect the ball on drawrod
A4
for damage.
4. Move the sliding tube
A3
on the jaw insert
A
to check for
ease of motion and the initial position of the blade
E2
.
5. Visually inspect the blade
E2
for damage.
6. Check the insulation of drawrod
A4
for damage.
7. Check the coating of jaw insert
A
for damage.
8. Check the seal ring
A2
for damage.
9. Check whether the drawrod
A4
or sliding tube
A3
is
bent.
Shank tube
1. Visually inspect the insulation for damage.
2. Check whether the shank tube is bent.
3. Check that the fluted knob
B2
can be turned easily.
4. Check the trigger
B3
for proper operation.
Handle
1. Check the ratchet lever
C1
, locking lever
C3
and
buttons
C4
for ease of motion.
2. Check cable connector
C2
for damage and corrosion.
MN031-447-S1 EN