15
Display of contact quality
Good
Sufficient
Insufficient
Not connected
EASY, BABY
MONO
3.5 |
SELECTING THE NEUTRAL ELECTRODE
Neutral electrode application
Select the application site for the neutral electrode in such
a way that the current paths between the active and neutral
electrodes are as short as possible and run lengthways or
diagonal to the body (as muscles boast higher conductivity in
the direction of the fibrils).
- EASY: For monitoring of split neutral electrodes
- BABY: For monitoring of split pediatric electrodes
- MONO: For selection of one-piece neutral electrodes
2
3
1
1
1
Display of neutral electrode
contact quality
2
Type of neutral electrode:
EASY, BABY, MONO
3
Selected neutral
electrode is
highlighted in white
When pediatric electrodes are
selected, the maximum power
of the monopolar current type is
reduced to 50 watts.
1. USING THE ARC 350
|
2. DISPLAY AND CONTROL ELEMENTS
|
3. START-UP
|
4. OPERATION
|
5. DISINFECTION AND CLEANING
|
6. TECHNICAL DATA