WARRANTY AND REPAIR
The Aaron 940™ is covered under warranty for a period of two years. The handpiece is covered under warranty
for a period of one year or 25 steam autoclave cycles, whichever comes first. The warranty becomes null and
void if damage occurs from incorrect handling or misuse of the product.
For warranty and repair work, please contact Bovie® and obtain a Return Goods Authorization number (RGA).
Place the number so that it can be seen on the exterior of the package and ship directly to Bovie. A return
without an RGA may not be accepted.
TROUBLESHOOTING
The Aaron 940™ has been designed and manufactured with the utmost safety in mind. The unit is equipped to
automatically detect a malfunction. The following table list error codes, their meaning and
recommended actions to be taken to resolve the error.
Error Code
Description of Error
Recommended Action
E1
Activation upon power up
• Check handpiece for activation.
• Check footswitch for activation; once the activation is
halted the unit will resolve the error. If the error
persists the handpiece could be malfunctioning and
may need to be replaced.
E2
DC Supply Over Voltage
• Switch unit off and on again.
Detection
• Make sure unit is connected to correct power source
for the unit.
E3
Pulse Width
• Switch unit off and on again.
E4
Delta Error
• Switch unit off and on again.
E5
Multiple Errors
• Switch unit off and on again.
E6
Temperature Error
• Switch unit off. Allow unit to cool. Switch unit on.
If problems persist, the unit should be taken out of service and the manufacturer should be notified. For
technical support or return authorization phone +1-800-537-2790.
TRANSPORT AND STORAGE
Ambient temperature range: -40˚C to 70˚C
Relative humidity: 10% to 100%, including condensation
Atmospheric pressure: 50kPa to 106kPa.
OUTPUT POWER CHARACTERISTICS
Maximum Output for Bipolar and Monopolar
Power readouts agree with actual power into rated load to within 20% or 5 watts, whichever is greater.
14
Bovie Medical Corporation
Mode
Output Power
Output Frequency
Repetition Rate
Vp-p max
Coagulation
40 W @ 1000
Ω
550 kHz
±
50 kHz
21 kHz
±
10%
8.0 kV
Bipolar
40 W @ 200
Ω
550 kHz
±
50 kHz
21 kHz
±
10%
2.5 kV
Содержание Aaron 940
Страница 1: ......
Страница 2: ...1 User s Guide Aaron 940 USER S GUIDE ...
Страница 18: ...17 Gebruikershandleiding Aaron 940 GEBRUIKERSHANDLEIDING ...
Страница 34: ...33 Guide de l utilisateur Aaron 940 33 GUIDE DE L UTILISATEUR ...
Страница 50: ...49 Bedienungsanleitung Aaron 940 49 BEDIENUNGSANLEITUNG ...
Страница 66: ...65 Guido do utilizado Aaron 940 65 GUIDA PER L UTENTE Guida per l utente Aaron 940 ...
Страница 82: ...81 Guía del usuario Aaron 940 GUÍA DEL USUARIO 81 ...
Страница 98: ...97 Guido do utilizado Aaron 940 GUIA DO UTILIZADOR ...
Страница 114: ...113 Käyttöopas Aaron 940 KÄYTTÖOPAS ...
Страница 130: ...129 Användarhandbok Aaron 940 ANVÄNDARHANDBOK ...
Страница 146: ......