Vercise™ DBS Programming Manual
91053785-17 33 of 1705
Programming the Patient
en
Understanding the Charge Density Warnings
The Clinician Programmer software provides several warnings when the charge density
at a contact may exceed 30 microcoulombs/cm
2
per stimulation phase, as there is
evidence of tissue damage above this level. The graph below displays the recommended
maximum charge density of 30 microCoulombs/cm
2
for different combinations of current
amplitude (mA) and pulse width (µs) for Boston Scientific DBS Leads.
30 uC/cm
2
Current Amplitude
(mA)
Pulsewidths (usec)
The Clinician Programmer software provides a warning when the charge density at a
contact is about to be programmed to exceed 30 microcoulombs/cm
2
per stimulation
phase. However, the Clinician Programmer allows the stimulation to continue above this
level. The warning reads:
“CAUTION: Charge density limit reached. Continue programming at your
discretion.”
Note:
A charge density limit warning will be displayed on the Programming screen
while any area is above the charge density limit and will clear when all areas are
below it.
A warning is also displayed when the amplitude range of a program exceeds charge
density limits. The warning reads:
“Caution: With present settings, the patient could exceed the charge density
limit with the remote control. Continue programming at your discretion.”
Another warning is displayed when a program that is capable of exceeding the charge
density limit is saved to a program slot in the patient’s Remote Control. The warning
reads:
“Caution: Charge density limit may be reached with the current Max mA set.
Do you want to continue?”
Содержание VERCISE DBS
Страница 257: ...fr Manuel de programmation du DBS Vercise 91053785 17 251 1705 Cette page a volontairement été laissée vierge ...
Страница 853: ...da Vercise programmeringshåndbog til dyb hjernestimulering DBS 91053785 17 847 af 1705 Denne side er tilsigtet tom ...
Страница 939: ...pt EU Manual de Programação do Vercise DBS 91053785 17 933 de 1705 Esta página foi deixada em branco intencionalmente ...
Страница 1264: ...Vercise DBS برمجة دليل 1705 من 1258 91053785 17 Vercise DBS برمجة دليل ar سلك إضافة مدعومة غير المخصصة الميزة هذه ...
Страница 1367: ...el Εγχειρίδιο προγραμματισμού DBS Vercise 91053785 17 1361 από 1705 Αυτή η σελίδα αφέθηκε σκόπιμα κενή ...
Страница 1453: ...cs Příručka pro programování systému DBS Vercise 91053785 17 1447 z 1705 Tato stránka je záměrně ponechána prázdná ...
Страница 1623: ...pl Podręcznik programowania Vercise DBS 91053785 17 1617 z 1705 Strona celowo pozostawiona pusta ...