Servicing

This screen is password protected and accessed by Boston 

Scientific personnel only.

PROPER CARE AND CLEANING
Overview

Proper care of the Rez

ū

m™ Generator is very simple, yet it is 

an important factor in its reliability. This section describes the 

proper care and cleaning required for the generator.

Warning:

 No modification of this equipment is allowed. Do not 

attempt to service or maintain the generator while in use with a 

patient.

Precaution:

 When any of the following occur, turn the power 

OFF, remove all accessories from the patient, and unplug the 

power cord from the outlet.
•  There is smoke or a strange odor leaking out of the 

generator.

•  The generator has been dropped or impacted by an object.
•  Liquid or foreign matter gets inside the generator.
•  Damage to the generator is suspected. 

Recommended Care

Warning:

 Failure on the part of all responsible individuals, 

hospitals, or institutions, employing the use of Rez

ū

Generator, to implement the recommended cleaning schedule 

may cause equipment failure and possible health hazards. 

The manufacturer does not, in any manner, assume the 

responsibility for performing the recommended maintenance 

schedule. The sole responsibility rests with the individuals, 

hospitals, or institutions utilizing the Rez

ū

m Generator.

Warning:

 Do not submerge the device in liquids or pour 

cleaning liquids over, into or onto the generator.

Warning:

 Before conducting proper care and cleaning, turn 

the power OFF and unplug the power cord from the outlet to 

prevent electric shock.

Precaution:

 To prevent damage to generator, do not clean 

any part of the device with phenolic compounds. Do not 

use abrasive or flammable cleaning agents. Do not steam, 

autoclave, or gas-sterilize the generator.

Precaution:

 Using this generator with the ventilation ports 

blocked could cause a breakdown. Clean this generator with 

care.

Precaution:

 Only approved equipment and accessories shall be 

connected to the generator.

Precaution:

 The Rez

ū

m Generator USB port is only intended for 

use by authorized service personnel or to export treatment data.
To ensure the Rez

ū

m Generator is always functional when  

required, Boston Scientific recommends performing the 

following routine activities:
•  Performing a Visual Inspection.
•  Cleaning the Rez

ū

m Generator.

•  Routine Care per checklist in this section.
It is important that the generator is stored at room temperature if 

it is expected to be used.
The Rez

ū

m Generator requires no calibration.

Visual Inspection

Warning:

 After the visual inspection, if the device is damaged 

or a message indicating the generator is not to be used, take 

the generator out of service and call Technical Assistance.
The generator should be carefully inspected prior to installation 

and use.
•  Carefully inspect the generator case for stress or physical 

damage.

•  Inspect all external connections for loose connectors.

•  Inspect the external power cord for damage or cracking.
•  Inspect the display for marks, scratches, or other damage.
•  Verify that the product label on the device is clearly legible 

and present.

Cleaning the Rez

ū

m Generator

It is recommended that the generator be inspected after each 

use according to the Routine Care Checklist in this manual and 

cleaned when appropriate. Listed below are recommendations 

for cleaning the generator. The generator does not need to be 

sterilized before or after use.

Warning:

 Do not submerge the device in liquids or pour 

cleaning liquids over, into or onto the generator.

Precaution:

 After cleaning, allow complete drying before 

plugging in to an outlet by wiping with a dry, soft cloth.

Precaution:

 Do not soak the generator or accessories in any 

medical liquid. Also, keep liquids out of the generator and 

accessories.

Precaution:

 When using disinfectant solutions, follow the 

manufacturer’s directions.

Precaution:

 Using this generator with the air vent blocked could 

cause a breakdown. Clean this generator with care.

Precaution:

 To prevent damage to equipment, do not clean 

any part of the generator with phenolic compounds. Do not 

use abrasive or flammable cleaning agents. Do not steam, 

autoclave, or gas-sterilize the generator.

Recommended Cleaning Products

The following cleaning products may be used to clean the 

exterior surfaces of the generator:
• Water
•  70% Isopropyl Alcohol
•  Super Sani-Cloth® Germicidal Disposable Wipes by PDI only

Not Recommended Cleaning Products

•  Do not use abrasive cleaners or strong solvents such as 

acetone or acetone-based cleaners.

•  Do not mix disinfecting solutions (such as bleach and 

ammonia) as hazardous gases may occur.

•  Do not clean electrical contacts or connectors with bleach.

Cleaning Instructions

1.  Before cleaning the generator, turn the generator OFF and 

disconnect the power cord.

2.  To remove any foreign material and fluid (e.g. dust, paper, 

etc.), wipe thoroughly with a soft cloth lightly dampened 

with water or 70% isopropyl alcohol. Super Sani-Cloth 

wipes may be used per manufacturer instructions. To 

prevent scratching the display, the use of a soft cloth is 

recommended.

3.  When cleaning, do not immerse.
4.  Wring any excess moisture from the cloth before and during 

cleaning.

5.  Avoid pouring fluids on the generator, and do not allow fluids 

to penetrate the exterior surfaces of the generator.

6.  To dry the generator after cleaning, wipe with a dry, soft cloth.

Routine Care Checklist

Routine Care activities involve verifying operation and safety.
Maintenance should always be performed by the customer 

at least once every 12 months. The following checklist is 

recommended to be utilized when checking the generator:

Table 6: Routine Care Checklist

Visual Inspection

Warning:

 After the visual inspection, if the generator is 

damaged or a message indicates to not use, take the 

generator out of service and call Boston Scientific Technical 

Assistance.

•  Carefully inspect the generator case for stress or 

physical damage.

•  Inspect all external connections for loose connectors. 
•  Inspect the external power cord for damage or cracking.
•  Inspect the display for marks, scratches, or other damage.
•  Verify that the product label on the device is clearly 

legible and present.

Operating Test

Warning:

 If a Critical Error message is displayed, take the 

generator out of service and call Technical Assistance.

•  Set-up generator and turn ON power to check the start-

up diagnostics.

Authorized Repair or Service

Warning:

 Shock Hazard - Do not open, disassemble, or alter 

the Rez

ū

m Generator. Failure to observe this warning can 

result in personal injury or death. Refer maintenance issues to 

authorized service personnel.

Warning:

 Do not use the generator if it is damaged, is not 

functioning properly, or fails to meet an electrical safety check. 

Notify the appropriate personnel to ensure the generator is 

removed from service and properly repaired. The generator 

has no user-serviceable internal components. Try to resolve 

any maintenance issues with the generator by using the 

Troubleshooting Table presented in Section 6, Troubleshooting. 

If you are unable to resolve the problem, contact Technical 

Assistance.  

 

The warranty will be void upon unauthorized disassembly or 

service of the Rez

ū

m Generator.

TECHNICAL SPECIFICATIONS
Overview

This section contains specifications for the Rez

ū

m Generator 

and EMC information.

Generator Specifications

Precaution:

 Exposing the Rez

ū

m Generator to extreme 

environmental conditions outside of normal parameters may 

compromise the ability of the Rez

ū

m Generator to function 

properly and/or cause the plastic to warp and/or crack.

Precaution:

 If the Rez

ū

m Generator is stored in an environment 

with a temperature below the room temperature, the unit should be 

allowed to warm up to the needed operating temperature before 

using.

Table 7: Generator Specification

Description

Specification

Protection Against 

Electric Shock

Class I equipment (generator)

Model Number

G2200-003

Voltage and Frequency

100-240 VAC, 50-60HZ

Power Input

10 amps maximum

External Fuses

Two, 10ah-250v, 5 x 20mm

Mode of Operation

Continuous operation

System Control

Provides controlled flow of water 

vapor at ambient temperatures 

below 25˚C

Case Dimensions

23L x 16W x 9H inches

Weight

50 pounds or less (generator only)

Power Cord Length

2.5 m to 3.05 m

Applied Parts 

Protection

 Type BF

Содержание rezum G2200-003

Страница 1: ...igure 17 Treatment in Progress Screen 6 Figure 18 Pending Ready Indicator on Screen 6 Turbo Flush 6 Connecting New Delivery Device During Treatment Session 6 Figure 19 New Delivery Device 6 Turning Off the Rezūm Generator 6 Figure 20 Power Button 6 Rezūm Generator Options Menu Items 6 Figure 21 Options Menu Button 6 Figure 22 Options Menu Choices 6 Drain Bladder 6 Figure 23 Drain Bladder Confirmat...

Страница 2: ...d adjacent to or stacked with other equipment If adjacent or stacked use is necessary test the generator to verify normal operation Refer to the Electromagnetic Immunity information Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm to any part of the Rezūm Generator including cables specified for use with the...

Страница 3: ...ntrols and Connections Warning The generator contains magnets in the LCD lid Avoid close or prolonged contact with electrical devices or devices that have strong magnetic fields Warning The Rezūm Generator is equipped with a USB port that is sensitive to ESD that may potentially result in injury or device failures Warning Do not connect a grounding wire from a grounding stud to a gas pipe or water...

Страница 4: ...r in the following locations Locations where gases and flames are used Locations where the air includes dust salt or sulfur Locations exposed to prolonged direct sunlight Locations that vibrate or are subject to sharp impacts Locations near heating equipment Locations where chemicals are stored This generator cannot be used in any room in which noise generating apparatuses are used such as an MRI ...

Страница 5: ...Delivery Device a progress bar will be initiated and displayed The text will change and a tone will be generated to indicate that the operation is in progress Priming is initiated by deploying the needle and holding the Delivery Device vapor activation button for the priming duration which lasts approximately 30 seconds A message will be displayed indicating when priming is complete If the vapor a...

Страница 6: ...e to continue with current procedure record Figure 19 New Delivery Device Turning Off the Rezūm Generator In the event of power failure the generator will automatically shut OFF Turn the power button OFF Please remove the Delivery Device from the patient immediately per instructions in the IFU Turn on again to restart the generator to begin a new therapy session Precaution Before moving this gener...

Страница 7: ...generator will automatically delete the oldest record to perform another therapy session The procedure records are saved in both a csv and txt format The serial number plus date time and unique number are saved as the file name The txt file contains all the user viewable information stored in the procedure record Figure 30 The csv file contains details of individual treatments stored in that proce...

Страница 8: ...tions and click OK Figure 37 Select Files to Export Figure 38 Reading Files Select export path and select Save Figure 39 Export Files After saving the files the generator will build the archive export the files and prompt when it is successful Figure 40 Building Archive Figure 41 Exporting File Figure 42 Export Successful Procedure Summary The Procedure Summary screen displays a sortable list of r...

Страница 9: ...quipment do not clean any part of the generator with phenolic compounds Do not use abrasive or flammable cleaning agents Do not steam autoclave or gas sterilize the generator Recommended Cleaning Products The following cleaning products may be used to clean the exterior surfaces of the generator Water 70 Isopropyl Alcohol Super Sani Cloth Germicidal Disposable Wipes by PDI only Not Recommended Cle...

Страница 10: ...mmon mode 100kHZ Mains power quality should be that of a typical commercial or hospital environment Surge immunity EN IEC 61000 4 5 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Power frequency magnetic field EN IEC 61000 4 8 30 A m 50 60 HZ Power frequency magnetic fields should be at levels characteristic of a typical l...

Страница 11: ...crowave treatment devices and there may be an impact on measurement precision for patients using cardiac pacemakers and other similar devices Check the operation of this generator during and after use of the above mentioned equipment and with patients potentially affected ESSENTIAL PERFORMANCE The Rezūm System essential performance is defined as Enable visualization of intraurethral anatomy and tr...

Страница 12: ...I Program Files Corrupted Power off on the Generator If problem persists contact Technical Support 35001 Generator Error Unexpected GUI Program Exit Power off on the Generator If problem persists contact Technical Support 35002 Generator Error Unable to Start GUI Program Power off on the Generator If problem persists contact Technical Support 40000 Generator Error GUI Unable to Communicate with MC...

Страница 13: ...5 Faulty Delivery Device Needle Deployment Error Ensure needle is retracted Replace Delivery Device 296 Faulty Delivery Device Needle Retraction Error Reattempt needle retraction If problem persists retract needle manually and replace Delivery Device 300 Saline Pump Error Saline Pump Encoder Error Ensure Delivery Device saline flush line is correctly inserted into saline pump and pump door is clos...

Страница 14: ...arranted Product or component furnished to address a warranty claim will be the longer of the unexpired term of the warranty applicable to the repaired or replaced Warranted Product or 90 days Boston Scientific warrants all Equipment that is loaned or otherwise remains the property of Boston Scientific as part of Boston Scientific s Agreement with the Customer Remedies If Customer promptly notifie...

Страница 15: ... AUS Do not use if package is damaged Argentina Local Contact ARG Para obtener información de contacto de Boston Scientific Argentina SA por favor acceda al link www bostonscientific com arg Catalog Number REF EU Authorized Representative EC REP blue safety sign Follow Instructions For Use Contents Recyclable Package Australian Sponsor Address AUS Do not use if package is damaged Argentina Local C...

Отзывы: