27
Q-TECH™ ProgrammEr: oPEraTIon
Ending a patient Session
To end an Online patient session and return the programmer to its Offline operation mode:
1.
Select the Main Menu button on the Navigation Bar. The Main Menu screen ap-
pears.
2.
Select the End Session button (Figure 16).
Figure 16:
Main Menu screen
3.
A warning message will be displayed to the user if any of the following conditions
exists:
•
Automatic Therapy is programmed to Off
•
Reference S-ECG has not been acquired
•
Automatic Setup or Optimization has not been completed. This message is
typically displayed following the implant procedure as Setup Optimization was not
performed (Figure 17).
4.
Select the Continue button to end the patient session and return to the Startup Menu
screen, or select Cancel to remain online and return to the Main Menu screen.
Содержание Q-Tech 3200
Страница 1: ...Q TECH PROGRAMMER A COMPONENT OF THE S ICD SYSTEM USER S MANUAL 3200 ...
Страница 6: ......
Страница 81: ......