23
Q-TECH™ ProgrammEr: oPEraTIon
Modes of Operation for the SQ-rX device
The SQ-RX device has three modes of operation:
•
Shelf
•
Therapy On
•
Therapy Off
Shelf Mode
The Shelf mode is a low power consumption state intended for storage only. When communication
is initiated between the SQ-RX device and the programmer, a full-energy capacitor reformation is
automatically performed and the SQ-RX device is prepared for set-up. Once the SQ-RX device is
taken out of Shelf mode, it cannot be reprogrammed back into Shelf mode.
Therapy On Mode
The Therapy On mode is the primary operating mode of the SQ-RX device, allowing automatic
detection of, and response to, ventricular tachyarrhythmias.
Note:
The SQ-RX device must be programmed out of Shelf mode before being
programmed to Therapy On.
Therapy Off Mode
The Therapy Off mode disables automatic therapy delivery and enables manual control of shock
delivery. Programmable parameters may be viewed and adjusted via the programmer. Also, the
subcutaneous electrogram (S-ECG) may be displayed or printed.
The SQ-RX device automatically defaults to Therapy Off mode when the SQ-RX device is taken out
of Shelf mode.
Note:
Manual and rescue shock therapy are available once the initial Setup process
is complete.