Vercise
™
DBS Remote Control 4 Directions for Use
Vercise™ DBS Remote Control 4 Directions for Use
en
MRI Mode
A patient implanted with an MR Conditional DBS System can undergo an MR Conditional Scan when all
instructions listed in the
ImageReady
TM
MRI Guidelines for Boston Scientific DBS Systems
are followed,
including enabling MRI Mode on the Stimulator. Contact Boston Scientific Technical Support or go to
www.bostonscientific.com/manuals for the latest version of this manual.
Enabling MRI Mode
When the Remote Control is linked to an MR Conditional Stimulator, the
Enter MRI Mode
icon will
appear on the
System Settings
screen. The Remote Control must be used to enable MRI Mode on the
Stimulator before performing an MRI scan on a patient. Once the MRI scan is complete, disable MRI
Mode. Do not leave the Stimulator in MRI Mode for extended periods of time beyond what is necessary to
perform the MRI scan. The stimulation is automatically turned OFF when MRI Mode is enabled.
Warning:
Do not perform an MRI scan if MRI Mode is not enabled. Scanning under different
conditions may result in patient injury or device malfunction.
To enable MRI Mode:
1.
Unlock the Remote Control by pressing the
Lock/Unlock
button on the right side of the Remote
Control
.
2. After unlocking the Remote Control, the
Home
screen appears.
Note:
The Remote Control may display either a text screen in one of the languages provided or an
iconic screen.
3.
Press the
Right Arrow
button to navigate to the
Main Menu.