Operator’s Manual
8-4
Table 8-1: System Specifications, Continued
Description
Specification
Back Panel Analog Outputs
Standard BNC connections to output
signals representing RF power,
impedance, and tip temperature.
Calibration Mode
Power Output
0 - 1 volt corresponds to 0 - 100 W,
continuous square wave
Impedance
0 - 1 volt corresponds to 0 - 300
Ω
,
continuous
square wave
Temperature
0 - 1 volt corresponds to 20 - 120
°
C,
continuous square wave