Controls and Displays
6-5
9.
If using the Footswitch, install its cable connector into the FOOTSWITCH Cable
Connector on the Controller rear panel (see Figure 6-5). Carefully push the cable
connector into the FOOTSWITCH connector until it is seated firmly in place.
"
Note
Do not twist or bend the FOOTSWITCH connector during insertion as this
action might cause damage to the pins.
10.
Position the Footswitch to allow easy access by the physician.
Figure 6-5: Connecting the Footswitch to the Controller Rear Panel
Ω
RF POWER
ANALOG
OUTPUT
SPEAKER
FOOT SWITCH
RS-232
MAIN
FUSES
WARNING
RISK OF FIRE.
REPLACE FUSE
AS MARKED.
WARNING
RISK OF FIRE.
REPLACE FUSE
AS MARKED.
OUTPUT POWER 150W
OPERATING FREQUENCY 500KHz
IMPEDANCE
TEMPERATURE
SPARE
VOLUME
CAUTION: DO NOT OPEN
ELECTRICAL SHOCK HAZARD
100-120V
3A 50/60 Hz
220-240V
1,6A 50/60 Hz
AUX
MAINS
T3A 250V
T1,6A 250V
2X
100-120V
1A 50/60 Hz
220-240V
500mA 50/60 Hz
T1A 250V
T500mA 250V
2X
CAUTION: THE TOTAL SYSTEM
CHASSIS RISK CURRENT SHOULD NOT
EXCEED 100
µµµµ
A
ATTENTION: LE COURANT DE TISQUE
TOTAL'ADMISSIBLE AU CHASSIS NE
DOIT PAS DEPASSER 100
µµµµ
A.
AUX
FUSES
Footswitch
Footswitch
Cable
Connector
11.
The System is now ready to record electrograms prior to and during delivery of RF
power.
Turning the Controller “ON”
1.
Turn the Controller ON by pressing the power switch (to the “1” position) located in
the power input module on the Controller rear panel (see Figure 6-6).
The Controller automatically initiates a self-test procedure which is indicated by the
illuminated front panel displays and continuous audio tone generated for
approximately 2 seconds. If no system malfunction is detected, the Controller
changes to the READY Mode.
"
Note
The connection between the XP APM and Controller is not required to
complete the initial self-test.