5. Advance the Lead to the desired location.
6. Once the Paddle Lead is at the appropriate vertebral level, use standard technique to identify Leads
and proceed to the instructions for Connecting to the OR Cable in the appropriate DFU for your SCS
system as listed in your
Reference Guide
.
Intraoperative Stimulation Testing
Note:
The following steps are for procedural reference only. For detailed stimulation testing procedures
and guidelines refer to the appropriate programming manual for your SCS system as listed in the
Reference Guide.
1. After linking the Clinician Programmer to the Trial Stimulator, check impedances to verify that
components are properly connected. Lead impedance is measured and displayed for each of the
IPG’s Contacts. Impedances, indicated by an x or an orange circle in your Programming Software, are
considered to be resultant from open or unconnected wires. Refer to your
Programming Manual
for
additional information.
2. Using test stimulation, enlist patient feedback to verify Lead placement and pain coverage.
Note:
If Lead repositioning is necessary, turn stimulation off before proceeding.
3. Repeat steps 1 and 2 if the Lead has been repositioned.
4. When the desired stimulation is achieved:
a) Turn the Trial Stimulator off.
b) Unlock each OR cable connector and disconnect from the Lead(s).
5. Record the Lead position by capturing a fluoroscopic image to be sure the Leads have not moved.
Retest if necessary.
6. Proceed to “Anchoring the Lead.”
Note:
Before
receiving a permanent SCS system, it is recommended that patients undergo a trial
procedure so that they can experience stimulation to evaluate if SCS is effective at treating their
chronic pain
.
Anchoring the Lead
Leads can be permanently anchored with a Suture Sleeve or with an anchor. Refer to the Directions for
Use for your Boston Scientific anchor as listed in your
Reference Guide
, or continue with the following
steps to anchor using a Suture Sleeve.
1. Place a Suture Sleeve over the Lead and down to the supraspinous ligament or deep fascial tissue.
Содержание Artisan
Страница 15: ...Surgical Leads Directions for Use 92395574 02 11 of 277 en This page intentionally left blank ...
Страница 57: ...Gebruiksaanwijzing voor chirurgische leads 92395574 02 53 van 277 nl Deze pagina is met opzet leeg gelaten ...
Страница 113: ...Οδηγίες χρήσης χειρουργικών απαγωγών 92395574 02 109 από 277 el Αυτή η σελίδα παρέμεινε σκόπιμα κενή ...
Страница 127: ...Bruksanvisning for kirurgiske elektroder 92395574 02 123 av 277 no Denne siden er blank med hensikt ...
Страница 141: ...Brugsanvisning til kirurgiske elektroder 92395574 02 137 af 277 da Denne side er med vilje efterladt tom ...
Страница 155: ...Kirurgisten johtimien käyttöohje 92395574 02 151 277 fi Tämä sivu on jätetty tarkoituksella tyhjäksi ...
Страница 169: ...Bruksanvisning till kirurgiska elektroder 92395574 02 165 av 277 sv Den här sidan har avsiktligt lämnats tom ...
Страница 183: ...Návod k použití chirurgických elektrod 92395574 02 179 z 277 cs Tato stránka je záměrně ponechána prázdná ...
Страница 197: ...Návod na použitie chirurgických elektród 92395574 02 193 z 277 sk Táto strana je zámerne ponechaná prázdna ...
Страница 211: ...Sebészeti Vezetékek használati utasítás 92395574 02 207 277 hu Ez az oldal szándékosan lett üresen hagyva ...
Страница 225: ...Instrukcja obsługi elektrod chirurgicznych 92395574 02 221 z 277 pl Strona celowo pozostawiona pusta ...
Страница 253: ...Upute za upotrebu kirurških elektroda 92395574 02 249 od 277 hr Ova je stranica namjerno ostavljena prazna ...
Страница 281: ...Surgical Leads Directions for Use 92395574 02 277 of 277 This page intentionally left blank ...