Instructions for use
Measure
English
English
English
English
English
English
English
1 8
PPB
N E W M E A S U R E M E N T
Unassigned Result
07.04.2016 11:43
A S S I G N T O P AT I E N T
NOTE
QC
The result “5 ppb -LO-” or “<5 ppb” represents a result below the lower detection
limit of the device. The result “300 ppb - HI- or “>300 ppb” represents a result
above the upper detection limit of the device.
When performing a measurement without selecting a patient before, the result is
shown as unassigned. You can assign the measurement to a patient immediately
or later from the list of measurements:
1. Select the
ASSIGN TO PATIENT
button.
The patient list is opened.
2. Select the dedicated patient from the list.
3. Measurements assigned to a patient can be unassigned by the
UNASSIGN
button.
Failed measurement
M E A S U R E M E N T FA I L E D
Please wait until the
clock
symbol
on the handheld display
has disappeared. Then press
the
Enter-button
to start a new
measurement.
In case your measurement failed, the
MEASUREMENT FAILED
screen opens.
1. The screen might show being busy to prepare the next measurement. Wait until
the clock icon on the handheld stops blinking. Press the
ENTER
button on the
handheld and the measurement procedure starts again with the
VISUALIZA-
TION
screen and the animation on the screen indicates that the regeneration
process is ongoing.
The display depends on the selected visualization, see
“Visualization and regeneration” in 4.1.