1 689 975 257
2019-05-15
|
Robert Bosch GmbH
30 | ACS 652 | Component test and maintenance
en
12.6 Vacuum pump
12.6.1 Changing vacuum pump oil
DANGER – Risk of electric shock from
dangerous voltage!
Electric shocks resulting from contact with
live parts (e.g. master switch, printed cir-
cuit boards) can cause injury, heart failure
and death.
¶
Disconnect the mains plug before open-
ing the ACS 652.
ATTENTION – Risk of burns from hot
surfaces
Contact with the hot vacuum pump surface
will cause severe burns.
¶
Allow the vacuum pump to cool down.
¶
Wear protective gloves.
i
The vacuum pump oil must always be changed after
10 hours of operation. The prompt
"
REPLACE OIL
"
appears on the screen when the vacuum pump oil
needs changing.
i
Use the vacuum pump oil (item number
SP00100086) as specified by Bosch.
1
3
2
Fig. 5:
Vacuum pump
1 Sight class
2 Oil fill plug
3 Drain plug
!
Do not use excessive force when securing and re-
moving the oil drain and oil filling screw.
i
The container for catching oil should have a capacity
of approx. 1 l.
1.
Place a vessel under the drain.
2.
Remove drain plug and oil fill plug from vacuum
pump.
3.
Drain off all the oil.
4.
Screw on drain plug finger-tight.
5.
Fill in vacuum pump oil.
6.
Start the vacuum phase.
7.
Check the oil level (sight glass).
i
The oil level should be in the middle between the
"full" and "empty" marks.
12.6.2 Reset of oil change interval
!
Observe the ACS 652 menu prompting.
1.
Call up start screen.
2.
Press <
3
> and <
Flush
> simultaneously and release.
The message
ENTER CODE 0000
is displayed.
3.
Enter code
5555
and confirm with
E
.
4.
Select "
CHANGE OIL
" with
o
u
and confirm with
E
.
Current runtime of vacuum pump is displayed.
5.
Press <
MENU
> until vacuum pump runtime is set to
zero.
"
Oil change interval is reset.
i
Return with <
Stop
>.