98404704A
6
TKEJD --
For the currently selected measurement channel
returns a status flag
and the time delay in seconds between the left edge of the display
window and the first waveform edge of either slope. Trigger delay can
be used to move the window with respect to the trigger to select or
exclude portions of a long string of pulses. This command works only
with TKAMEAS active. After returning data the instrument returns to
the previous Talk mode, normally TKAMEAS.
Examples: send TKAMEAS send only once
send TKEJD “send TKEJD read” may be repeated
read 1, 1.163e-7 valid edge delay in seconds
or read 0, 0 no valid result
TKMMODE --
Returns a measurement mode run/stop flag, mode identifier and units
flag. For the run/stop flag 0 = STOP, 1 = RUN. For the mode
identifier 0 = Pulse, 1 = CW
,
2 = CDF, 3 = 1-CDF, 4 = PDF. For the
units flag 0 = log (dBm) and 1 = linear (watts). After returning data
the instrument returns to the previous Talk mode.
Examples: send TKMMODE
read 1, 0, 0 running in pulse mode with log units
or read 0, 3, 1 stopped in 1-CDF mode with linear
units
1.6 EOI Only Talk mode Terminator
In the
Util > IEEE-488 > Bus Setup > Talk Term
menu a new choice,
EOI only
, has been added.
This allows return strings to be terminated only by the EOI signal of the GPIB, simplifying setup
with controllers which use this as their default.
1.7 Reference Lines in Linear Units mode
Reference Lines and Reference Line Tracking now work in the Linear as well as Logarithmic
units modes. Reference line level readout is always in dBm.
2. CORRECTIONS
2.1 Sensor Temperature Readings
All previous versions and revisions of Model 4400/4500/4400A and 4500A report sensor
temperature approximately 10 degrees Celsius lower than the actual internal sensor temperature.
This characteristic has been of little consequence since only delta temperature values are used.