PRODUCT INFORMATION 

 
The  Boekel  series  of  Pipet  Dryer  and  Combination 
Washer/Dryer  are  manufactured  from  top  quality 
stainless  steel  and  are  designed  to  provide 
automatic  washing  and  heat  drying  of  glass  pipets 
up  to  375mm  long.    The  series  consists  of  a  Pipet 
Dryer  only  unit  and  a  Pipet Washer/Dryer  unit  that 
combines both the washing and drying functions. 
 
Model  137100  &  137100-2  that  provide  washing 
capability  can  be  hooked  directly  to  the  lab  faucet 
(with user supplied hose) or permanently connected 
to  the  lab’s  water  supply.    This  model  is  designed 
with  a  natural  “air  gap”  in  the  incoming  water  fill 
tube  to  prevent  back  siphoning  into  the  water 
supply. 
 
The  models  that  provide  electrical  heat  drying 
capabilities are available in both 115 VAC and 230 
VAC versions. 
 

UNPACKING 

 
1.  Remove the unit from the carton. 
 
2.  Inspect  for  damage.    Report  all  shipping 

damage  to  the  carrier  immediately.    Shipping 
damages  are  covered by  the  carrier  and  repair 
and/or  replacement  for  shipping  damage  must 
be coordinated through the carrier. 

 
3.  Retain  all  packaging  material  in  the  event  that 

the unit must be returned. 

 
4.  Completely fill out and return the self addressed 

Boekel  Warranty  Registration  card  to  Boekel 
Scientific. 

 

SETUP AND OPERATION OF PIPET DRYER 
(Models 1372 and 1372-2) 

 
1.  Place the unit on a level surface and insert the 

power  cord  plug  into  a  grounded  receptacle  of 
appropriate voltage. 

 
 
 

 

2.  For efficient operation, begin Pipet Dryer warm 

up while pipets are being washed.  To start the 
Pipet Dryer, 

move the toggle switch to the “ON” 

position.  The red pilot light should illuminate.   

 

3.  Allow  wet  pipets  to  drain  before  placing  them 

into a hot dryer.  Place pipets tip down into the 
Pipet  Dryer.    An  optional  pipet  holder  (model 
13741/2) is available for transferring pipets and 
holding pipets in the dryer. 

 

4.  Drying  time  will  vary,  depending  on  quantity 

and size of glassware.  The approximate drying 
time is one hour. 

 

5.  NEVER allow the Pipet Dryer to run overnight.  

High  temperature  can  build  up  and  cause  the 
heater to burn out. 

 

SETUP 

AND 

OPERATION 

OF 

PIPET 

WASHER/DRYER (Model 137100 & 137100-2) 

 

1.  Place  the  unit  on  a  level  surface.    A  level 

surface  is  required  for  the  unit  to  operate  and 
drain properly.  If a slanted drain board is used, 
the unit must be built up on the low side so that 
the unit sits level. 

 
2.  A  1/2"  I.P.S.  hose  barb  is  supplied  for 

connecting  the  unit  to  a  lab  faucet  with  a  user 
supplied rubber hose.  This hose barb may be 
removed 

for 

alternate 

forms 

of 

water 

connection.    NOTE:  Some  states  require  the 
use  of  an  approved  vacuum  breaker  in  the 
water  supply  line.    Please  consult  your  local 
plumbing codes. 

 

3. 

A 1” I.P.S. thread is provided on the drain port 
to allow permanent connection of a drain line if 
desired.    Any  such  drain  line  must  be  capable 
of  handling  the  three  gallons  of  water  that  are 
instantly  released  during  washing  action.    The 
drain must not back up during water release. 

 
 