Boekel 137100 Скачать руководство пользователя страница 2

4. 

Typical  procedure  for  cleaning  pipets:

  Load 

the pipets into the pipet holder supplied with the 
unit.  Then, using a glass jar, soak the pipets in 
a  Dichromatic  acid  solution  up  to  50%  in 
strength.    Allow  pipets  to  drain  when  removing 
them  from  the  Dichromatic  solution,  and  then 
place  them  immediately  into  the  Washer  or 
Washer/Dryer  unit  and  turn  on  the  water  to 
initiate the washing action. 

  
5.  Care should be taken to minimize the amount of 

acid  solution  that  drips  into  the  Washer  or 
Washer/Dryer  as  acids  can  have  a  detrimental 
effect on the units. 

 
6.  The  complete  washing  cycle  of  filling  the  unit 

with  water  and  draining  is  accomplished  in 
approximately  70  seconds,  depending  on  the 
incoming  water  flow.    It  is  recommended  that 
water  flow  be  set  at  about  2  to  3  gallons  per 
minute for good washing action.  Ten to fifteen 
washing 

cycles 

with 

hot 

water 

are 

recommended  for  best  washing  action.    Cold 
water however can be used. 

 
7. 

To initiate the drying cycle on Washer/Dryer 
models  137100  and  137100-2:

  insure  that  the 

water  for  the  washing  cycle  has  been  shut  off 
and  that  the  water  has  fully  drained  from  the 
unit. 

 
8.  The  wet  pipets  from  the  washing  cycle  should 

be  drained  outside  the  Pipet  Washer/Dryer 
before beginning their drying.  Pipets should be 
in tip down for best drying results. 

 
9. 

Insure that the unit’s power cord plug has been 
inserted  into  a  properly  grounded  outlet  of 
appropriate voltage. 

 
10. Activate  the  dryer  heater  by  moving  the  toggle 

switch to the “ON” position.  The red pilot light 
should illuminate. 

 
11. Drying time will vary, depending on quantity and 

size of glassware.  The approximate drying time 
is one hour. 

 
12. When drying is 

complete, switch the unit “OFF” 

and allow the unit to cool for 30 minutes before 
beginning the next washing cycle. 

 
 

 

13. NEVER allow the drying cycle to run over night.  

High  temperature  can  build  up  and  cause  the 
heater  to  burn  out  and  cause  possible  solder 
gaps resulting in leaks. 

 

CLEANING 

 
All  Pipet  Washer/Dryer  and  Dryer  models  can  be 
internally  rinsed  with  clean  water  and  wiped 
externally  with  a damp  cloth.    The  Pipet  Dryer  and 
Washer/Dryer  models  must  be  unplugged  before 
any cleaning is performed.  The pipet holding rack, 
when used,  should always  be  rinsed  if  it  was  used 
in acidic solution. 
 

SERVICE: 

RETURN 

AUTHORIZATION 

IS 

REQUIRED 

 
Should 

service 

be 

required, 

contact 

your 

salesperson  or  call  Boekel  Scientific  at  215-396-
8200 or 800-336-6929. 
 

SPECIFICATIONS 

 

PIPET  WASHER/DRYER  Models  137100  and 
137100-2 

Height: 23 3/4” 
Cylinder Diameter: 6” 
Base Diameter: 10 1/2” 
Model 137100 uses 115 VAC, 195 Watts 
Model 137100-2 uses 230 VAC, 195 Watts 
Includes 1374 1/2  Pipet Holder 
 
 

PIPET DRYER Model 1372 and 1372-2 

Height: 23 1/2” 
Cylinder Diameter: 5 1/2” 
Base Diameter: 10 1/2” 
Model 1372 uses 115 VAC, 275 Watts 
Model 1372-2 uses 230 VAC, 275 Watts 
Optional 1374 1/2  Pipet Holder 

 