9
7.
User Menu
The initial display of the User Menu will show the most relevant
information, including battery level, day counter, currently set
volume, time, activated sensors (see chapter 8) and Bluetooth. The
dots at the bottom of the screen are to indicate the current menu
page.
On the second page, the Pen Counter is shown. To reset this pen
counter, press and hold the left display button for two (2) seconds.
On the third page, the Day Counter is shown. To reset the day
counter, press and hold the left display button for two (2) seconds.
On the fourth page, the volume can be adjusted. The currently set
volume is shown in the middle, the optional volume is shown on the
lower left side. Press the left display button to choose the optional
volume. The dose will be changed only upon usage of the device (after
the rinsing cycles and before the first injection).
The fifth page shows the Service Counter, which is counting from
12,000 to 0. If the value falls below 0, the counter continues counting
negatively. Press and hold the left display button for two (2) seconds
to reset the service counter.
A ‘Service due’ message appears after there have been 12,000 or
more injections, showing that a cylinder service needs to be
performed (see chapter 11.1)
The sixth page shows the Total Counter. This counter cannot be reset.
The last page shows the submenu “Settings”. Enter the submenu with
the left display button. Pushing the right display button returns the
screen to the initial display (see above).
Submenu “Settings”:
On the first page of the submenu, the Position Sensor can be
controlled. Press the left display button to disable or enable the
position sensor. Recommended setting is an enabled Position Sensor.
(Disabling is only recommended when using the tubing adapter
supplied with the device and having vaccine delivered via tubing.) For
further details, see chapter 8 “Sensors”.