Troubleshooting Guide                                     

 

                  BM-300 

1. 

Optical System

 

TROUBLE 

CAUSE 

SOLUTION 

The  nosepiece  is  not  in  the  located  position 
(objective and light path not coaxial) 

Locate the nosepiece properly 

where it clicks 

The surface of the lamp becomes black 

Change a new lamp bulb 

1.The  edge  of  the  field  of 
view is dark or the brightness 
is not uniform 

A lens (the objective, condenser, eyepiece or 
collector) is dirty. 

Clean it thoroughly 

Dirt/dust on the specimen 

Replace with a clean specimen 

2.  Dirt  or  dust  is  visible  in 
the field of view 

Dirt/dust on the eyepieces 

Clean them 

Specimen is not covered 

Add cover glass on it 

The  thickness  of  the  cover  glass  is  not 
suitable 

Use standard cover glass with 

thickness of 0.17mm 

Specimen is placed reversely 

Turn it over 

Dry  objective  has  oil  on  it.  (especially  for   
40X objectives) 

Wipe the oil 

A lens (the objective, condenser, eyepiece or 
collector) is dirty. 

Clean it 

Immersion  oil  is  not  used  with  the  100x 
objective 

Use specified oil 

Air bubbles existed in the immersion oil 

Eliminate the bubble 

The aperture iris diaphragm is stopped down 
too far 

Adjust the aperture iris 

diaphragm properly 

3.Visibility is poor 
Image is not sharp; 
Contrast is poor; 
Details are indistinct 

Dirt or dust on the eyepiece 

Clean it 

Condenser is not properly centered 

Center the condenser with the 

centering screw 

The nosepiece is not properly engaged 

Engage the nosepiece properly 

 
 

4.One  side  of  image  is 
blurred 

 

The specimen is not clamped 

Clamp it with stage clips 

5.The image is yellowish 

The blue filter is not used 

Use the blue filter 

The aperture iris diaphragm is too small 

Adjust it properly 

The condenser is too low 

Adjust it properly 

6.The  brightness  is  not 
enough 

 
 

A lens (the objective, condenser, eyepiece or 
collector) is dirty. 

Clean it 

 

 

 

Содержание BM-300

Страница 1: ...his manual is for biological microscope Model BM 300 To ensure the safety obtain optimum performance and to familiarize yourself fully with the use of this microscope it is recommended strongly that y...

Страница 2: ...R NOTICES 2 1 Components Name 3 2 Assembly 4 2 1 Assembly Diagram 2 2 Assembly Procedures 4 5 3 Adjustment 8 3 1 Adjustment Set Diaphragm 8 3 2 Adjustment Procedures 11 4 Technical Specifications 13 5...

Страница 3: ...ngeover switch 8 Always use the power cord provided by Novel II Maintenance and Care 1 All the lenses have been adjusted properly do not dismount them by yourself please 2 The nosepiece and coarse and...

Страница 4: ...3 1 Components Name BM 300 Eyepiece Binocular viewing head Nosepiece Objective Main body...

Страница 5: ...nents The number in the figure shows the assembly steps Before installing be sure every components is clean do not score any parts or glass surface Keep well with hexagon wrench provided When changing...

Страница 6: ...ing head Fig 1 2 Insert the binocular viewing head into the microscope head and turn it to a proper position then screw down the bolt to fix it 2 2 2 Installing the eyepiece Fig3 4 Insert the eyepiece...

Страница 7: ...agnification Installing objective this way will make the change of magnification to be easier during using Clean the objective regularly for lens is susceptible to dust When operating use 10 magnifica...

Страница 8: ...power supply receptacle Make sure the connection is well Do use the supplied power cord all the time If lost or damaged select the same standard cord please A wide range of voltage like 100V 240V is...

Страница 9: ...peration BM 300 3 1 Adjustment set diagram Fig 14 15 Right Fine Focus Knob Right Coarse Focus Knob Tension Adjustment Collar Main Switch Portrait Adjustment Knob Lateral Adjustment Knob Brightness Adj...

Страница 10: ...9 BM 300 Left Coarse Focus Knob Interpupillar Distance Indicator Fig 12 Clocking Set Condenser Focusing Knob Left Fine Focus Knob Aperture Iris Diaphragm Collar Indicator...

Страница 11: ...center of the mechanical stage and use the stage clips to clamp it 2 Turn the portrait and lateral adjustment knob of the mechanical ruler move the specimen to the required position Be careful when ch...

Страница 12: ...n one The condenser and the objective are coaxial It has been adjusted before leaving factory so the user needn t to adjust them by self The highest position of the condenser has been adjusted too It...

Страница 13: ...viewing head is the diopter in use 3 3 8 Adjusting the tension adjustment collar Fig 23 Turn the tension adjustment collar with your fingers When the collar is turned in the direction of the arrow te...

Страница 14: ...00 Focusing Coaxial Coarse and Fine Adjustment Moving Range20mm Fine Division 0 002mm Condenser Abbe Condenser NA 1 25 Stage Double Layers Mechanical Stage 140mm 140mm Moving Range 75 50mm Illuminatio...

Страница 15: ...ead 1 Condenser Abbe Condenser NA 1 25 1 Nosepiece Quadruple 1 3W LED 1 Illumination Spare BGX 1 20 0 25A Fuse 2 2 Eyepiece 10 Wide Field Plan Eyepiece 18 mm 2 10 Wide Field Plan Eyepiece 20 mm 2 Infi...

Страница 16: ...t especially for 40X objectives Wipe the oil A lens the objective condenser eyepiece or collector is dirty Clean it Immersion oil is not used with the 100x objective Use specified oil Air bubbles exis...

Страница 17: ...at of the other Interpupillary distance is incorrect Adjust interpupillary distance The diopter is not proper Adjust the diopter properly 5 Observation is tiring The brightness of the illumination is...

Отзывы: